Sonel MIC-10 Operating Manual
Text of the manual
INSULATION RESISTANCE METER
MIC-10
SONEL Test & Measurement
Santa Clara, Ca. USA
SONEL SA
Świdnica, Poland
Version 1.1 17/03/2017
Thank you for purchasing the MIC-10 insulation tester. Please acquaint yourself with this manual to
avoid measuring errors and prevent possible problems related to the operation of the meter.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) th
OPERATING MANUAL MIC-10 version 1.1
3
CONTENTS
1
SAFETY .................................................................................................................... 5
2
METER CONFIGURATION .................................................................................. 6
3
MEASUREMENTS ...............................................................
OPERATING MANUAL MIC-10 version 1.1
5
1 Safety
The MIC-10 meter is designed to determine the safety of electrical wiring, to ensure adequate
protection against electric shock. For correct operation and accurate results observe the following
recommendations:
Before operating the meter acquaint yourself thoroughly with this manual. Observe the safety
cauti
OPERATING MANUAL MIC-10 version 1.1
6
2 Meter Configuration
Turn on the meter by pressing the
power and SET/SEL buttons
simultaneously.
Use the
and
buttons to select a
parameter.
Use the
and
buttons to change a
parameter value. The value or symbol to be
changed flashes.
The
symbol indicates an active parameter,
the
- symbol indicates an inactive one.
Folo
OPERATING MANUAL MIC-10 version 1.1
7
3 Measurements
3.1 Measurement of insulation resistance
WARNING:
Measured object must be de-energized. Do not make measurements on live
circuits.
Important Note:
Make sure that test leads do not touch each other and the probes (crocodile clips)
during measurements, especially of high resistances Such contact may cause
th
OPERATING MANUAL MIC-10 version 1.1
8
The meter is ready
for measurement.
Press and hold START button.
The measurement is performed continuously
until the START button is released or the pre-
set time is reached.
Instead of holding the START button down
the function can be latched by first pressing
the ENTER and then press START and
release. The
symbol wil