EX570-ENG V2.3 3/09
2
Introduction
Congratulations on your purchase of the Extech EX570 True RMS Autoranging Multimeter.
This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency,
Duty Cycle, Diode Test, and Continuity plus Thermocouple Temperature and InfraRed
Temperature. It features a waterproof, rugged design for heavy duty use. Proper use and
care of this meter will provide many years of reliable service.
Safety
This symb
EX570-ENG V2.3 3/09 3
CAUTIONS
•
Improper use of this meter can cause damage, shock, injury or death. Read and
understand this user manual before operating the meter.
•
Always remove the test leads before replacing the battery or fuses.
•
Inspect the condition of the test leads and the meter itself for any damage before
operating the meter.
•
Use great care when making measurements if the voltages are greater than 25 VAC
rms or 35 VDC. These vol
EX570-ENG V2.3 3/09 4
SAFETY INSTRUCTIONS
This meter has been designed for safe use, but must be operated with caution. The rules
listed below must be carefully followed for safe operation.
1.
NEVER apply voltage or current to the meter that exceeds the specified maximum:
Input Protection Limits
Function
Maximum Input
V DC or V AC
1000 VDC/AC rms
mA AC/DC
500 mA 1000V fast acting fuse
A AC/DC
10A 1000V fast acting fuse (20A for 30 seconds
max every 15 m
EX570-ENG V2.3 3/09 5
Controls and Jacks
1. IR Thermometer
2. Laser Pointer
3. 40,000 count LCD
4. MAX/MIN button
5. MEMORY button
6. SELECT button
7. Function switch
8. mA, µA and 10A input jacks
9. PEAK and RELATIVE button
10. RANGE button
11. HOLD and
(Backlight) button
12. IR Thermometer button
13. Positive input jack
14. COM input jack
Note: Tilt stand and battery compartment are on rear of unit.
Symbols and Enunciators
•)))
Continuity
Diode t
EX570-ENG V2.3 3/09 6
Operating Instructions
WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very
dangerous and should be measured with great care.
1. ALWAYS turn the function switch to the OFF position when the meter is not in use.
2. If “OL” appears in the display during a measurement, the value exceeds the range you
have selected. Change to a higher range.
AC/DC VOLTAGE MEASUREMENTS
CAUTION: Do not measure DC voltages
EX570-ENG V2.3 3/09 7
AC/DC CURRENT MEASUREMENTS
CAUTION: Do not make 20A current measurements for longer than 30 seconds.
Exceeding 30 seconds may cause damage to the meter and/or the test leads.
1. Insert black test lead banana plug into the negative COM jack.
2. For current measurements up to 4000µA DC, set the function switch to
the µA position and insert the red test lead banana plug into the µA/mA
jack.
3. For current measurements up to 400 mA