The AFA1000/E must be field-calibrated once the

room air supply and exhaust is proportionally balanced. When the unit is powered up, the following sequence of events occurs: 1. The alarm performs a self-test of its functions, LEDs and audible alarm (approximately 3 seconds). 2. At the end of the delay, the unit will do one of two things: a. If the ...

Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK

T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com 8. If the sample is accepted, lower the sash by half and measure the face velocity using a calibrated instrument. Using the + / - buttons enter the measured face ...

Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK

T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com Control Settings The ECON CONFIG menu includes the following control parameters:- 1. MANUAL / AUTO -- Allows the control output to be set to Auto or Manual cont ...

Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK

T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com ...

Control Panel Dimensions

Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com Control Panel Cutout Dimensions (NOT shown to scale) 5.83” (148mm) 3.46” (88mm) 1.57” (40mm) 1.57” (40mm) 4.65” (118mm) 3.15” ...

Temperature Electronics Ltd. Unit 2, Wren Nest Road, Glossop SK13 8HB UK

T. +44 (0) 1457 865 635 F. +44 (0) 1457 868 843 E. [email protected] w. www.tel-uk.com Airflow Sensor Installation Diagram For complete manual and product information, log on to www.tel-uk.com ...

TEL AFA1000/E MK2 user guide recommended for: AFA1000/2 MK2, AFA4000/1/AS, Modulex E8, NanoMatic, AFA1000/BLD.