Hioki 3274 Instruction Manual
Read Online
Read the Hioki 3274 Instruction Manual online using the built-in PDF viewer.
Text of the manual
CLAMP ON PROBE
3274
Instruction Manual
May 2015 Revised edition 5
3275A981-05 15-05H
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Notes on Safety
1
Notes on Use
4
Chapter 1 Overview
9
1.1 Product Overview
9
1.2 Features
9
1.3 Names of Parts
10
1.4 Parts of the Sensor
11
Chapter 2 Specifications
13
2.1 Product Specifications
13
2.2 Standards Applying
15
Chapter 3 Measurement Procedure
19
3.1 Notes on Use
19
3.2 Preparations for Measurement
20
3.3 Demagnetizing and Zero Adjustment
21
3.4 Measurement Procedure
22
3275A981-05
This device is designed to comply with IEC 61010
Safety Standards, and has been thoroughly tested for
safety prior to shipment. However, mishandling
during use could result in injury or death, as well as
damage to the device. Be certain that you understand
the instructions and precautions in the manual before
use. We disclaim any responsibility for accidents or
injuries not resulting directly from device defects.
Introduction
Inspection
Notes on Safety
Thank you for purchasing this HIOKI 3274, 3275 CLAMP ON
PROBE.
To obtain maximum performance from the device, please read this
manual first, and keep it handy for future reference.
When you receive the device, inspect it carefully to ensure that no
damage occurred during shipping. If damage is evident, or if it
fails to operate according to the specifications, contact your dealer
or Hioki representative.
Supplied accessories
Instruction manual
1
Carrying case
1
symbol printed on the device indicates that the user
should refer to a corresponding topic in the manual
(marked with the
symbol) before using the relevant
function.
In the manual, the
symbol indicates particularly
important information that the user should read before
using the device.
Indicates that only insulated conductors suited to the
voltage of the circuit under test can be measured.
DANGER
Indicates that incorrect operation presents an extreme
hazard that could result in serious injury or death to the
user.
WARNING
Indicates that incorrect operation presents a significant
hazard that could result in serious injury or death to the
user.
CAUTION
Indicates that incorrect operation presents a possibility of
injury to the user or damage to the device.
NOTE
Indicates advisory items related to performance or correct
operation of the device.
Safety Symbols
This manual contains information and warnings essential for safe
operation of the device and for maintaining it in safe operating
condition. Before using the device, be sure to carefully read the
following safety notes.
The following symbols in this manual indicate the relative
importance of cautions and warnings.
Do not measure around a bare conductor. Doing so
may result in short-circuit or electric shock. Take
measurements at a location on an insulated wire
where there is sufficient insulation for the circuit
voltage.
Refer to the derating characteristics when
measuring current that includes a highfrequency
component and never measure any current that
exceeds the rated current.
Use with high frequencies or strong magnetic fields
may cause the device to become abnormally hot,
resulting in fire, equipment damage, or burns. (See
〝2 Specifications〟(p. 12).)
Observe the following to avoid electric shock and
short circuits.
Connect the device to the 3269 or the 3272 Power
Supply and waveform measurement
instrument (oscilloscope or recorder) first,and then
to the active lines to be measured.
When the sensor is opened, do not shortcircuit the
conductor being measured or other two wires with
the metal part of the tip.
Be careful to avoid damaging the insulation surface
while taking measurements.
Notes on Use
Follow these precautions to ensure safe operation and to obtain
the full benefits of the various functions.
Before clamping the conductor being measured,
make sure that the insulation on the conductor is
undamaged. Also, take care not to damage the
insulation when clamping the conductor. Any damage
to the insulation could cause an electric shock.
This instrument is made for use with the 3269 or
3272 POWER SUPPLY.
To prevent fire or damage of the measurement
target and device as well as burns, exercise
caution concerning the following when measuring
high-frequency currents or currents that contain
high-frequency components:
Eddy current loss may cause heating of the
sensor head.
Dielectric heating may cause heating of cord
insulation and other materials.
This device should only be connected to the
secondary side of a breaker, so the breaker can
prevent an accident if a short circuit occurs.
Connections should never be made to the primary
side of a breaker, because unrestricted current flow
could cause a serious accident if a short circuit
occurs.
Be sure to observe all operating precautions for the
waveform monitoring instrument (oscilloscope or
recorder) and other measurement instruments to
which this device is connected.
When using a measurement instrument that does
not provide isolation between its input terminals
and chassis or other input terminals, please pay
attention to the following points.
If a signal is applied to an input terminal other than
that to which this device is connected, do not
connect the ground-side terminal to any non-
ground potential. Otherwise, short-circuit current
will flow through the 3269 or 3272, or this device
from the ground terminal, which could cause an
electrical accident or damage.
Power Source
Load
Load
Do not allow the device to get wet, and do not take
measurements with wet hands. This may cause an
electric shock.
Do not press the demagnetizing switch (DEMAG) to
perform demagnetization while the conductor being
measured is clamped. Doing so could damage the
circuitry or cause an accident that might result in
injury or death.
Ensure that the input does not exceed the
maximum rated current to avoid device damage,
shortcircuiting and electric shock resulting from
heat building.
CAUTION
To avoid damage to the device, protect it from vibration
or shock during transport and handling, and be
especially careful to avoid dropping.
Do not store or use the device where it could be
exposed to direct sunlight, high temperature, humidity,
or condensation. Under such conditions, the device
may be damaged and insulation may deteriorate so
that it no longer meets specifications.
Before using the device the first time, verify that it
operates normally to ensure that the no damage
occurred during storage or shipping. If you find any
damage, contact your dealer or Hioki representative.
This device is not designed to be entirely water- or
dust- proof. To avoid damage, do not use it in a wet or
dusty environment.
The sensor head is a precision assembly including a
molded component, a ferrite core, and a Hall effect
element. It may be damaged if subjected to sudden
changes in ambient temperature, or mechanical strain
or shock, and therefore great care should be exercised
in handling it.
The matching surfaces of the sensor head are
precision ground, and should be treated with care. If
these surfaces are scratched, performance may be
impaired.
Do not apply a static electricity or other source of high
voltage to the sensor. Doing so may damage its
internal Hall elements and circuitry.
Foreign substances such as dust on the contact
surfaces of the sensor head can cause acoustic
resonance and degrade measurement, so it should be
cleaned by gently wiping with a soft cloth.
Refer to NOTE (p26) for resonance sound.
To avoid damaging the sensor cable and power supply
cable, do not bend or pull the cables.
When the power is on, keep closed, except when
clamping them onto the conductor to be measured.
The facing surface of the core section can be scratched
while it is open.
Do not short-circuit the output terminal and do not input
voltage to the output terminal. The device may be
damaged.
Note that the device may be damaged if the applied
current exceeds the measurement range.
When the 3269, 3272 Power Supply or waveform
measurement instrumentʼs power is turned off, do not
apply current to the device. Doing so may damage the
device, the 3269, 3272 Power Supply and/or waveform
measurement instrument.
NOTE
Correct measurement may be impossible in the presence of strong
magnetic fields, such as near transformers and high-current
conductors, or in the presence of strong electromagnetic fields
such as near radio transmitters.
Service / Maintenance
To clean the device, wipe it gently with a soft cloth moistened
with water or mild detergent. Never use solvents such as
benzene, alcohol, acetone, ether, ketones, thinners or gasoline, as
they can deform and discolor the case.
If the device seems to be malfunctioning, contact your dealer or
Hioki representative.
When sending the device for repair, pack carefully to prevent
damage in transit. Include cushioning material so the device
cannot move within the package. Be sure to include details of
the problem.
Hioki cannot be responsible for damage that occurs during
shipment.
Use the original packing materials when transporting the device,
if possible.
Transport the device in its carrying case.
Chapter 1 Overview
___________________________________________________________________
Chapter 1
Overview
1.1 Product Overview
1.2 Features
This device can be directly connected to a BNC input connector
of a waveform measuring instrument such as an oscilloscope or
recorder, and by clamping on a conductor to be measured, allows
the current waveform to be easily captured.
Highly accurate current detection.
Easy current measurement.
Broadband frequency characteristics.
Model 3274 : DC to 10 MHz
Model 3275 : DC to 2 MHz
Large diameter allows high-current measurements.
Easy protection function to avoid self-heating during excessive
input.
Unique HIOKI development of thin film Hall effect element
Chapter 1 Overview
___________________________________________________________________
Current direction indication
"LOCK" indication
Terminator
Sensor cable
Power supply cable
"UNLOCK" indication
(inside)
Sensor
7
8
1
4
3
2
5
6
2
1.3 Names of Parts
External view
Chapter 1 Overview
___________________________________________________________________
1.4 Parts of the Sensor
1. Clamp
This clamps around the conductor to be measured.
2. Slider
This slider opens the clamp. Always use it to open and close the
clamp.
3. Lever
This lock mechanism keeps the clamp closed.
4. Sensor head
This clamps the conductor being measured, and carries out the
actual current measurement. It is a precision assembly including a
molded component, a ferrite core, and a Hall effect element. It
may be damaged if subjected to sudden changes in ambient
temperature, or mechanical strain or shock, and therefore great
care should be exercised in handling it.
5. Demagnetizing switch (DEMAG)
This demagnetizes the core if it has been magnetized by switching
the power on and off, or by an excessive input. Always carry out
demagnetizing before measurement.
The demagnetizing process takes about 3 second. During
demagnetizing, a demagnetizing waveform is output.
6. Zero adjustment dial (ZERO ADJ)
Use the zero adjustment dial to correct for the effect of a voltage
offset or temperature drift on the device.
When beginning measurement, after demagnetizing always carry
out zero adjustment.
Chapter 1 Overview
___________________________________________________________________
NOTE
7. Output connector
The current waveform of the measured conductor is output at a
constant rate (0.01 V/A).
Connect to the BNC input connector of the waveform measuring
instrument.
The output of this device is terminated internally.
Since the output resistance is approx. 7 Ω (model 3274) and
approx. 40 Ω (model 3275), the device must be used with a
high-impedance input to the measuring instrument. With an input
impedance of 50 Ω, accurate measurement is not possible.
If using BNC-banana plug adapters or similar to connect to input
terminals other than BNC connectors, make sure the polarity is
correct.
Turn the collar until it clicks, and check that it is locked
securely.
8. Power plug
Connect this to the 3269 or 3272 POWER SUPPLY receptacle to
supply power to the sensor terminator.
Chapter 2 Specifications
___________________________________________________________________
Bandwidth
Model 3274 : DC to 10 MHz (-3 dB)
(Typical characteristics shown in Fig.1)
Model 3275 : DC to 2 MHz (-3 dB)
(Typical characteristics shown in Fig.2)
Rise time
Model 3274 : 35 ns or less
Model 3275 : 175 ns or less
Maximum
continuous input
range
Model 3274 : 150 A
(Derating according to frequency shown in Fig.3)
Model 3275 : 500 A
(Derating according to frequency shown in Fig.4)
Maximum peak
current value
Model 3274 : Non-continuous 300 A peak
500 A peak at pulse width ≦30μs
Model 3275 : Non-continuous 700 A peak
Output voltage rate
0.01 V/A
Amplitude
accuracy
Model 3274 : To 150 A:
1.0% rdg.
1 mV
150 A to 300 Apeak:
2.0% rdg.
(DC, and 45 to 66 Hz)
Model 3275 : To 500 A:
1.0% rdg.
5 mV
To 700 Apeak:
2.0% rdg.
(DC, and 45 to 66 Hz)
Noise
Equivalent to 25 mA rms or less (for 20 MHz band
measuring instrument)
Input impedance
Model 3274 (Typical characteristics shown in Fig.5)
Model 3275 (Typical characteristics shown in Fig.6)
Chapter 2
Specifications
2.1 Product Specifications
Guaranteed at 23
±3
(73
±5.4
) after the power has
been on for 30 minutes.
Chapter 2 Specifications
___________________________________________________________________
Temperature
coefficient for
sensitivity
2% or less
Model 3274 (Input: 55 Hz 150 A, within a range of 0
to 40
, within a range of 32 to 104
)
Model 3275 (Input: 50 Hz 500 A, within a range of 0
to 40
, within a range of 32 to 104
)
Maximum rated
power
Model 3274 : 5.5 VA
Model 3275 : 7.2 VA
(within maximum continuous input range)
Rated supply
voltage
Model 3274 :
12 V
1 V
Model 3275 :
12 V
0.5 V
Operating
temperature and
humidity range
0 to 40
(32 to 104
), 80 % RH or less (no
condensation)
Storage
temperature and
humidity range
-10 to 50
(14 to 122
), 80 % RH or less (no
condensation)
Location for use
Indoor, altitude up to 2000 m (6562 feet), Pollution
Degree 2
Period of
guaranteed
accuracy
1 year (Opening/Closing up to 10,000 times)
Effect of external
magnetic fields
Model 3274 : Equivalent to a maximum of 150 mA
Model 3275 : Equivalent to a maximum of 800 mA
(in a DC or 60 Hz, 400 A/m magnetic field)
Diameter of
measurable
conductors
20 mm dia.
0.79" dia.
Measurable
conductors
Insulated conductor
Cable lengths
Sensor cable Approx. 2 m (78.7")
Power supply cable Approx. 1 m (39.4")
External
dimensions
Sensor
Approx. 176W X 69H X 27D mm
Approx. 6.93"W X 2.72"H X 1.06"D
Terminator Approx. 27H X 55W X 18D mm
Approx.1.06"W X 2.17"H X 0.71"D
Mass
Model 3274 : Approx. 500 g (Approx. 17.6 oz.)
Model 3275 : Approx. 520 g (Approx. 18.3 oz.)
Accessories
Instruction manual, Carrying case
Chapter 2 Specifications
___________________________________________________________________
Safety
EN61010
EMC
EN61326
2.2 Standards Applying
Chapter 2 Specifications
___________________________________________________________________
Fig.1 Frequency characteristics (Typical) (3274)
Fig.2 Frequency characteristics (Typical) (3275)
Fig.3 Derating according to frequency(3274)
Chapter 2 Specifications
___________________________________________________________________
Fig.6 Input impedance (Typical) (3275)
Fig.4 Derating according to frequency(3275)
Fig.5 Input impedance (Typical) (3274)
Chapter 2 Specifications
___________________________________________________________________
Chapter 3 Measurement Procedure
___________________________________________________________________
Chapter 3
Measurement Procedure
3.1 Notes on Use
Before using the instrument, make sure to refer to〝Notes on Use
〟(p.3 to p.8).
Chapter 3 Measurement Procedure
___________________________________________________________________
CAUTION
Before turning the device on, make sure the source
voltage matches that indicated on the rear panel of the
3269 or 3272. Connection to an improper supply voltage
may damage the 3269 or 3272 and present an electrical
hazard.
3.2 Preparations for Measurement
(1) Have the 3269 or 3272 POWER SUPPLY, and oscilloscope or
recorder for waveform measurement ready.
(2) Turn the power switch off and connect the power cord.
(3) Connect the power plug of the 3274 or 3275 to the power
receptacle of the 3269 or 3272.
(4) Turn the 3269 or 3272 power switch on, and check that the front
panel power indicator lights.
(5) Wait at least 30 minutes after turning on the device.
Immediately after power is supplied, offset drift may increase due
to the effects of self-heating of the device and other factors.
To ensure accurate measurement, wait at least 30 minutes after
turning on the device before performing measurement.
Chapter 3 Measurement Procedure
___________________________________________________________________
CAUTION
When disconnecting the output connector, be sure to
release the lock before pulling off the connector. Forcibly
pulling the connector without releasing the lock, or pulling
on the cable, can damage the terminator.
If using BNC-banana plug adapters or similar to connect
to input terminals other than BNC connectors, make sure
the polarity is correct.
Do not demagnetize while the 3274 or 3275 is clamping a
conductor to be measured. Demagnetizing causes current
to flow into the conductor, which may damage parts in the
circuit to be measured.
Demagnetizing
Conductor
Check that the conductor being measured is not clamped
when supplying power to the 3274 or 3275 for the same
reason. Demagnetized waveforms are generated when
supplying electric power.
3.3 Demagnetizing and Zero Adjustment
(1) With the waveform measurement instrument input at ground,
adjust the trace to the zero position.
(2) Set the input coupling of the waveform measurement instrument
to DC.
(3) Connect the output connector of the 3274 or 3275 to the input
connector of the waveform measurement instrument. Turn the
collar until it clicks, and check that it is locked securely.
Chapter 3 Measurement Procedure
___________________________________________________________________
I
3.4 Measurement Procedure
(4) Before clamping a conductor, confirm that the clamp can be
securely closed: press the slider until UNLOCK is no longer
displayed, and hold it until LOCK appears.
(5) Press the demagnetizing switch (DEMAG) on the terminator.
(6) Turn the zero adjustment dial on the terminator, to adjust the trace
to the zero position.
(1) Check that the system is safe, and that the preparations described
in the preceding section have been carried out.
(2) Pull the sensor slider, so that the clamp opens.
(3) Align the sensor so that the current direction indication
corresponds to the direction of current flow through the conductor
to be measured, and clamp so that the conductor is in the center
of the sensor aperture.
(4) Press the slider on the sensor head until the "UNLOCK"
indication disappears, and hold it until LOCK appears, and check
that the opening lever is firmly locked and the clamp securely
closed.
(5) It is now possible to monitor the current waveform. The output
rate of the 3274 or 3275 is 0.01 V/A. The current sensitivity can
be derived from the voltage sensitivity of the waveform
measurement instrument. For example, if the voltage sensitivity is
10 mV/division, the current sensitivity is 1 A/division.
Chapter 3 Measurement Procedure
___________________________________________________________________
NOTE
・ When using the 3275, note that two clamp-on probes may not be
used simultaneously with the 3272 POWER SUPPLY, depending
on the current to be measured.
・ The current consumption of clamp-on probes depends on the
current to be measured. Confirm that the total current
consumption of the clamp-on probes does not exceed the rated
output current of the 3272. See Figure 1.
Fig.7 Current consumption* vs. current to be measured (typical)(3274)
*The sum total of a positive and negative current consumption
Current consumption* vs. current to be measured (typical)(3275)
*The sum total of a positive and negative current consumption
Chapter 3 Measurement Procedure
___________________________________________________________________
CAUTION
The maximum continuous input range is based on
heat that is internally generated during
measurement. Never input current in excess of this
level. Exceeding the rated level may result in
damage to the probe.
The device may sustain damage from self-heating
even at current levels that are lower than the
maximum current value defined by the maximum
rated current.
The maximum rated current is a recommended
value that assumes sine-wave input under
standard conditions. Self-heating may increase if
the ambient temperature increases or the
measurement current waveform contains other
frequency components. Refer to the derating
characteristics in the product specifications.
If excess current is input, generated heat activates
a built-in safety function that blocks normal output.
If this happens, remove the input immediately
(remove the sensor from the conductor being
measured, or reduce the input current to zero).
Wait until the sensor has had sufficient time to cool
before resuming operation.
Heating generated during measurement of currents
with a frequency of 1 kHz or higher is mainly
attributed to the self-heating of the sensor heads.
In this case, the built-in safety function will not be
not activated. Be careful not to have accidents
including a burn by heat, short-circuit, and damage
to the sensor.
Even if the input current does not exceed the rated
continuous maximum, continuous input for an
extended period of time may result in activation of
the safety circuit to prevent damage resulting from
heating of the sensor.
At high ambient temperatures, the built-in safety
circuit may activate at current input levels below
the rated continuous maximum.
Chapter 3 Measurement Procedure
___________________________________________________________________
CAUTION
Continuous input of current exceeding the rated
maximum or repeated activation of the safety
function may result in damage to the device.
The product specifications that indicate the maximum
input range include a maximum peak current value in
addition to the maximum rated current.
The maximum peak current specifications are as
follows:
For Model 3274
300 A peak (discontinuous) … … … . (1)
500 A peak (with a pulse width of 30 μs or shorter)
… … … . (2)
For Model 3275
700 A peak (discontinuous) … … … . (3)
The values (1) and (3) determine the upper limits of
the response to waveforms. Use the instrument to
measure currents that do not exceed the maximum
continuous input range.
The value (2) determines the upper limit of the
response to single pulse inputs. Use the instrument
to measure currents that do not exceed the range
described above.
When opening the clamp, be sure to operate with
the slider.
Do not place any unclamped conductor with an
electric current of a frequency of 10 kHz or more
near the sensor head. Current flowing in the
conductor nearby may heat up the sensor head and
cause its temperature to rise, leading to damage to
the sensor. For example, when one side of a go-and-
return conductor is clamped and the other side is
also placed near the sensor head as shown in the
diagram, even if the electric current is lower than the
consecutive maximum current, electric currents in
both sides will heat up the wires and raise the
temperature, thereby causing damage to the sensor.
Chapter 3 Measurement Procedure
___________________________________________________________________
NOTE
The output of this device is terminated internally. Since the output
resistance is approx.7 Ω (3274) or approx.40 Ω (3275), the
must be used with a waveform measurement instrument that has an
input impedance of at least. Use a waveform measurement
instrument with an input impedance of at least 1 MΩ.
Immediately after powering on, this device may be subject to an
appreciable offset drift due to the effect of self-heating. To
counteract this, allow the device to warm up for about 30
minutes before carrying out measurement.
When performing continuous measurements, it is necessary to be
aware that the offset voltage drifts, depending on factors such as
the ambient temperature.
Under certain circumstances, oscillation may occur if the probe is
connected to the 3269 or 3272 POWER SUPPLY while the
power supply is on. This does not indicate a malfunction.
Oscillation can be stopped and operation restored to normal by
opening and closing the clamp.
Depending on the amplitude and frequency of the current being
measured, the sensor head may emit a resonant sound. This
sound may also occur during demagnetizing operation, but it does
not represent a malfunction (device failure).
If foreign matter becomes adhered to the facing surfaces on the
sensor head so that a slight gap exists between the upper and
lower sensors, the sensor head may emit a resonant sound. Any
foreign matter should be removed using the cleaning method
described in this manual (see〝Notes on use〟(p.3 to p.8)).
An increase in the volume of the resonant sound during use may
indicate that the gap between the upper and lower sensors has
increased in size. Since the sensor characteristics may change, it
is recommended to calibrate the device.
Acoustic resonance may occur depending on the level and
frequency of the measured current. This does not normally affect
measurements unless a foreign substance such as dust is present
on the contact surfaces of the sensor head.
Chapter 3 Measurement Procedure
___________________________________________________________________
Power source
Load
High
Low
Pressing the demagnetizing switch (DEMAG) will cause a
demagnetized waveform to be output from the instrument.
Although it may be asymmetry with respect to the zero-volt line,
the instrument has no malfunction.
The reading may be affected by the position within the clamp
aperture of the conductor being measured. The conductor should
be in the center of the clamp aperture.
When carrying out measurement, press the slider on the sensor
head until the "UNLOCK" indication disappears, and hold it until
LOCK appears, and check that the opening lever is firmly locked
and the clamp securely closed. Correct measurements cannot be
performed unless the clamp is securely closed and the slider is
pressed until LOCK is displayed.
At high frequencies, common mode noise may affect measurements
taken on the high voltage side of circuits. If this occurs, reduce the
frequency range of the waveform measuring instrument, or clamp onto
the low-voltage side of the circuit, as appropriate.
When power is turned on, a demagnetizing waveform is initially
applied to the output: this is intentional in the design, and not a fault.
Accurate measurement may be impossible in locations subject to
strong external magnetic fields, such as transformers and high-
current conductors, or in locations subject to strong external
electric fields, such as radio transmission equipment.
Chapter 3 Measurement Procedure
___________________________________________________________________
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com