SD Biosensor SD LipidoCare Manual
Text of SD Biosensor SD LipidoCare Manual
TABLE OF CONTENTS
CHAPTER 1. UNDERSTANDING THE NEW ANALYZER....................................... 4
1.
Indication for use; purpose of the analyzer .................................................................................. 4
2.
Test principle............................................................................................................................................... 5
3.
Before you start a test .....................................
Important Safety Instructions
Before using the SD LipidoCare Professional Analyzer, please read the following information.
Important Safety Instructions
The lancing device are for single patient use. Do not share them with anyone including other
family members! Do not use on multiple patients!
All parts of the kit are considered biohazardous and can potentially transmit infectious diseases,
even after you have performed cleaning and disinfection.
The Food
CHAPTER 1. UNDERSTANDING THE NEW ANALYZER
1. Indication for use; purpose of the analyzer
The SD LipidoCare Professional Analyzer is intended to be used by individuals in the
Professional to measure total cholesterol (TC), triglycerides (TG), HDL cholesterol (HDL), and
glucose in capillary whole blood with the SD LipidoCare Professional Lipid Profile Test Strips and
SD LipidoCare Professional Blood Glucose Test Strips; it is also intended to calculate low d
2. Test principle
LIPID
The SD LipidoCare Professional System combines enzymatic methodology and solid-phase
technology to measure total cholesterol (TC), HDL cholesterol (HDL), and triglycerides (TG).
Blood sample used for testing is whole blood from a fingerstick. The sample is applied to an SD
LipidoCare Professional Test Strip. The test strip is then placed into the SD LipidoCare
Professional Analyzer, where a unique system on the test strip separates
3. Before you start a test
CAUTION
Carefully read and follow the instructions and package inserts for the lipid/glucose test strips
and control solutions. It is very important to follow the instructions in order to prevent an
incorrect result or improper treatment.
Samples
For measuring lipid (cholesterol)
The SD LipidoCare Professional Analyzer for lipid testing is designed for testing a fresh capillary
whole blood sample from the fingertip.
For measuring
4. The SD LipidoCare Professional Analyzer
The Analyzer package includes following:
SD LipidoCare Professional Analyzer
SD Lipid Check Strip
SD Glucose Check Strip
User Manual
Quick Guide
4 AAA 1.5 V batteries
NOTE:
The SD LipidoCare Professional Lipid Profile Test Strips, SD LipidoCare Professional Blood
Glucose Test Strips, SD Lipid Control Solution, and SD Glucose Control Solution are required,
but not included, and must be purch
Provides access to the battery compartment
(4 AAA 1.5 V alkaline batteries)
H
Codechip slot
Insert a codechip for lipid testing
I
Optic cover
Holds the inserted test strip to prevent the movement of the test strip
when you are testing
J
Data port
Download test results to a personal computer
K
Printer port
Print test results on an external thermal printer
L
SET/PRT Button
Press to set the analyzer or print the test results
- 8 -
Display of the SD LipidoCare Professional Analyzer
The symbols in the display have the following meanings:
Indicates a codechip sign
Warning when the battery is low
Indicates printing
Indicates the index number of the
memory
Codechip number or memory
number stored in the memory
Appears during setting mode
Indicates measurement being
performed
Indicates beep sound setting
Indicates the operating
temperature is out of the range
Indicates Lipid Profile test