Scilogex D-500 User Instruction Manual
Text of Scilogex D-500 User Instruction Manual
CE/ cULus-Declaration of Conformity
EC-Declaration of Conformity
Unpacking the instrument
5
Proper use of the D-500
6
Dispersing Aggregates
7
Cleaning
8
Sterilization
9
Corrosion
9
Working with D-500
9
Working with Speed Control
9
Overload protection
10
Maintenance of Motor
10
Specification and Technical D-500
11
Stand
11
Assembly of H Stand
11
Assembly of the PT plate
The D-500 Disperser is a high speed dispersing instrument. It is based on the
Rotor/Stator Technology. The shaft and rotor/stator can be disassembled for easy
cleaning. You will receive the dispersing tools completely assembled for immediate use
with your disperser. Should you dismantle these (e.g. for cleaning) please refer to page
6. Th
• It is recommended that the operator should be using hearing protection when
operating the unit at maximum speed.
• Do not operate after the equipment malfunctions or has been damaged in any
manner. Return unit to our service center for examination and repair.
• Switch off the unit before changing the dispersing element.
• Only suitable
The voltage on the name plate must match the main voltage. If it does not, do not
operate the instrument.
Screw the support rod into the flange of the drive and tighten it. Make sure that the
“nylon” washer is between the flange and the nut to avoid scratching and rubbing
marks on the flange.
Make sure that the unit is switched off when connec
• Disassembly of the dispersing tool
1. Hold the dispersing tool with a cloth (to avoid cutting of your hand due to sharp
edges at the stator). At the same time turn the shaft tube clockwise. Screw the shaft
tube of the stator and pull it over the axle and put it aside.
2. Now hold the rotor with a cloth and turn the axle anticlockwis