Time TT130 Instructions Manual

Text of Time TT130 Instructions Manual

1. GENERAL DESCRIPTION ................................................................................................... 3

1.1 SCOPE OF APPLICATION........................................................................................................ 4 1.2 BASIC WORKING PRINCIPLE.................................................................................................. 4 1.3 BASIC CONFIGURATION AND EACH PART DESCRIPTION OF THE TESTER ..........

10.10 PROTECTIVE SHEATH OF PROBE...................................................................................... 17

11. PRECAUTIONS................................................................................................................... 18 11.1 AUTOMATIC TURN OFF..................................................................................................... 18 11.2 CLEANING OF TEST BLOCKS .............................................

1.1 Scope of application

Adopting the theory of ultrasonic wave measurement, TIME TT130 intelligent Ultrasonic Thickness Gauge possesses the capability of thickness measurement for various kinds of materials in which ultrasonic wave pulses propagate with a constant velocity and reflect on the back interface. The device can be used to perform accurate measurement for various kinds of plates and parts. Another prominent characteristic of it is to monitor v

LCD screen display:

BATT : low voltage indication 凸 : coupling indication m/s : metric unit of sound velocity mm : metric unit of thickness inch/μ s: imperial unit of sound velocity inch : imperial unit of thickness Keyboard:

ZERO : calibration

VEL : sound velocity ▲ : adjustment key for sound velocity, thickness and thickness unit ▼ : adjustment key for sound velocity, thickness and thickness unit VEL+ZERO : keys for thickness memory 2. Performance parameters Display type: 4-digit LCD Minimum display unit: 0.01 mm Working frequency: 5 MHz Measuring range: 1.2 mm~225.0 mm (steel) Lower limit for steel pipes: Ф 20 mm × 3.0 mm Measuring error: ± (1%

3. Main functions

1) Automatic calibration to zero: automatically correct the system errors 2) Automatic non-linear compensation: within the full range, computer software is used to correct the non-linear errors of the probe for the purpose of improving the accuracy 3) The upward and downward adjustment keys enable prompt selection of sound velocity, thickness, and check the thickness memory units 4) Prompt for coupling state: provide coupling indicatio

Sound velocity of last turnoff

4.2 The adjustment of sound velocity If the screen currently displays the thickness values, then push the VEL-key to enter into sound velocity state. The screen will display the content of the current memory unit of sound velocity. Each time the VEL-key is pushed the sound velocity memory unit will change. Five different sound velocity values will be displayed in turn. If the current displayed sound velocity is desired to

Put the coupling agent on the testing block supplied with the unit and couple the

probe with the testing block. At this time, the bar lines displayed on the screen will disappear one after the other until the screen display 4.0mm (0.158inch). This indicates that the calibration is over. Then enter the measurement state and measure a random test block. If the measuring value error exceeds the measurement error range, the calibration should be operated aga

5. Switch between meter and inch

5.1 When the power is off, press and hold ZERO button, press ON to toggle between meter and inch. 5.2 When the power is on, if the current display is inch, press ZERO for more than 5 seconds to toggle between meter and inch. Release the button to end setting. If you hit the ZERO quickly, the calibrate mode is on. 6. Sound velocity measurement If the sound velocity in a certain kind of material is to be measured, a test b