Time TIME2110 Manual 

Text of Time TIME2110 Manual 

Ultrasonic Thickness Gauge

TIME2110 (V20140815) Beijing TIME High Technology Ltd. 2 Content 1. GENERAL DESCRIPTION ............................................................................................................................................. 1 2. PERFORMANCE PARAMETERS ................................................................................................................................. 4

1. General description

1.1 Scope of application Adopting the theory of ultrasonic wave measurement, TIME2110 Ultrasonic Thickness Gauge possesses the capability of thickness measurement for various kinds of materials in which ultrasonic wave pulses propagate with a constant velocity and reflect on the back interface. The device can be used to perform accurate measurement for various kinds of plates and parts. Another prominent characteristic of it is to

LCD screen display:

BATT : low voltage indication 凸 : coupling indication m/s : metric unit of sound velocity mm : metric unit of thickness inch/μs: imperial unit of sound velocity inch : imperial unit of thickness Keyboard: ON : power ZERO : calibration VEL: sound velocity -: adjustment key for sound velocity, thickness and thickness unit : adjustment key for sound velocity, thickness and thickness unit VEL+ZERO : keys for thickness me

3) Automatic non-linear compensation: within the full range, computer software is used to

correct the non-linear errors of the probe for the purpose of improving the accuracy 4) The upward and downward adjustment keys enable prompt selection of sound velocity, thickness, and check the thickness memory units 5) Prompt for coupling state: provide coupling indication and the observation of the stability of the indication can tell if the coupling is normal 6

Sound velocity of last turnoff

Note: 5900 m/s= 0.232 inch/μs 5.2 The adjustment of sound velocity If the screen currently displays the thickness values, then push the VEL-key to enter into sound velocity state. The screen will display the content of the current memory unit of sound velocity. Each time the VEL-key is pushed the sound velocity memory unit will change. Five different sound velocity values will be displayed in turn. If the current displayed

At this time, the bar lines displayed on the screen will disappear one after the other until the screen

display 4.0mm (0.158inch). This indicates that the calibration is over. Then enter the measurement state and measure a random test block. If the measuring value error exceeds the measurement error range, the calibration should be operated again until the measuring value lies within the measurement error. 10 5.4 Measurement of thickness: Put the couplin

Time Measuring Instruments Owners Instructions: