Time TIME2110 Manual
Text of Time TIME2110 Manual
Ultrasonic Thickness Gauge
TIME2110
(V20140815)
Beijing TIME High Technology Ltd.
2
Content
1. GENERAL DESCRIPTION ............................................................................................................................................. 1
2. PERFORMANCE PARAMETERS ................................................................................................................................. 4
1. General description
1.1 Scope of application
Adopting the theory of ultrasonic wave measurement, TIME2110 Ultrasonic Thickness Gauge
possesses the capability of thickness measurement for various kinds of materials in which
ultrasonic wave pulses propagate with a constant velocity and reflect on the back interface.
The device can be used to perform accurate measurement for various kinds of plates and parts.
Another prominent characteristic of it is to
LCD screen display:
BATT : low voltage indication
凸
: coupling indication
m/s
: metric unit of sound velocity
mm
: metric unit of thickness
inch/μs: imperial unit of sound velocity
inch
: imperial unit of thickness
Keyboard:
ON : power
ZERO : calibration
VEL: sound velocity
-: adjustment key for sound velocity, thickness
and thickness unit
: adjustment key for sound velocity, thickness
and thickness unit
VEL+ZERO : keys for thickness me
3) Automatic non-linear compensation: within the full range, computer software is used to
correct the non-linear errors of the probe for the purpose of improving the accuracy
4) The upward and downward adjustment keys enable prompt selection of sound velocity,
thickness, and check the thickness memory units
5) Prompt for coupling state: provide coupling indication and the observation of the stability
of the indication can tell if the coupling is normal
6
Sound velocity of last turnoff
Note: 5900 m/s= 0.232 inch/μs
5.2 The adjustment of sound velocity
If the screen currently displays the thickness values, then push the VEL-key to enter into sound
velocity state. The screen will display the content of the current memory unit of sound velocity.
Each time the VEL-key is pushed the sound velocity memory unit will change. Five different
sound velocity values will be displayed in turn. If the current displayed
At this time, the bar lines displayed on the screen will disappear one after the other until the screen
display 4.0mm (0.158inch). This indicates that the calibration is over. Then enter the measurement
state and measure a random test block. If the measuring value error exceeds the measurement error
range, the calibration should be operated again until the measuring value lies within the
measurement error.
10
5.4 Measurement of thickness:
Put the couplin