CFM/CMM Thermo Anemometer +

InfraRed Thermometer Model AN200 UNITS AREA RECALL cf m/cmm Anemometer Thermometer AN200 NEXT PROBE TEMPERATURE www.GlobalTestSupply.com Find Quality Products Online at: [email protected]

Congratulations on your purchase of the Extech AN200 CFM/CMM Thermo Anemometer with

InfraRed Thermometer. This instrument measures Air Velocity, Air Flow (volume), Air Temperature (with probe) and Surface Temperature (with the InfraRed function). The large, easy- to-read backlit LCD includes primary and secondary displays plus numerous status indicators. The InfraRed feature includes a laser pointer for convenient tar

Power ON/OFF button

2. Probe input jack 3. Laser pointer 4. IR Sensor 5. Rubber holster 6. LCD Display 7. IR thermometer measurement button 8. Airflow buttons (4) 9. Air Temperature function buttons (2) 10. Vane 11. Airflow Average button 12. Backlight button 13. MAX‐MIN button for TEMPERATURE mode 14. HOLD for TEMPERTAURE functions 15. MAX‐MIN button for AIR VELOCITY/AIR FLOW (also used as�

Extech Instruments AN200 • Press to turn the meter ON or OFF

• IR + Laser Pointer Press and hold to measure. • MAX/MIN (Airflow) Record and store the highest and lowest airflow or velocity readings. ◄ (LEFT) also serves as change decimal point button in AREA mode • UNITS Press to select the mode of operation. In FLOW mode, the meter displays air volume. In VELOCITY mode, the meter displays air speed. ▲ (UP) also serves as inc

MAX (top of LCD): Max Hold function engaged for the Air Temperature function

 HOLD (top of LCD): Data Hold function engaged for the Air Temperature function  PROBE TEMP: Reminder that the top LCD digits represent Air (Vane) Temperature  : Indicates that the laser pointer is on.  IR TEMP: Indicates that the larger LCD digits represent IR temperature measurement  VEL: indicates that meter i

Extech Instruments AN200 Connecting the Vane

1. The vane plug is inserted in the meter’s sensor jack at the top of the meter. The plug and jack are keyed so that the plug can only fit in the jack one way. 2. Turn the plug carefully until it lines up with the jack and then firmly push the plug in place. Do not apply undue force or try to twist the plug side-to-side. 3. If the vane is not connected to the meter or if the sensor is defective, the

Air Flow Measurements (CMM / CFM)

1. Turn on the meter using the ON/OFF button 2. Press the UNITS button to select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet per minute). NOTE: At power up the meter will display the last unit of measure previously entered. 3. To begin entering the area in m 2 or ft 2, press and hold the AREA button until it beeps twice. The leftmost digit of the bottom displ

Extech Instruments AN200 Data Hold (Air Velocity/Air Flow)

1. While taking measurements, press the HOLD button to freeze the air velocity/air flow reading for later viewing. 2. The HOLD indicator will appear in the bottom of the LCD display. 3. Press HOLD again to return to normal operation. MAX/MIN/AVG Record (Air Velocity/Air Flow) This allows the user to record and view the highest (MAX), lowest (MIN) and average (AVG) readings. 1. Press the Airflow MA

InfraRed (Non-Contact) Temperature Measurements

1. The IR sensor is located at the top of the meter. 2. Point the sensor toward the surface to be measured. 3. Press and hold the red IR button to begin measuring the surface temperature of a desired target. IR TEMP and will appear on the display. The laser pointer will switch on to help aim the meter. 4. The measured IR surface temperature will appear at the center of the LCD

Extech Instruments AN200 m/s (meters per sec)

0.40 - 30.00 m/s 0.01 m/s ± (3%rdg + 0.20 m/s) km/h (kilometers/hour) 1.4 - 108.0 km/h 0.1 km/h ± (3%rdg + 0.8 km/hr) ft/min (feet per minute) 80 – 5900 ft/min 1 ft/min ± (3%rdg + 40 ft/m) mph (miles per hour) 0.9 – 67.0 mph 0.1 mph ± (3%rdg + 0.4 MPH) knots (nautical MPH) 0.8 to 58.0 knots 0.1 knots ± (3%rdg + 0.4 knots) Air Flow Range Resolution