FMS EMGZ309 Operating Manual
Text of FMS EMGZ309 Operating Manual
Digital Microprocessor Controlled Tension Amplifier
Version 2.12
08/2017 NS
Firmware Version
V2.10
Diese Bedienungsanleitung ist auch in Deutsch erhältlich.
Bitte kontaktieren Sie Ihren nächstgelegen FMS Vertreter.
© by FMS Force Measuring Systems AG, CH-8154 Oberglatt – All rights reserved.
Operation Manual EMGZ309
07.08.2017
2
Table of contents
1
Safety Instructions ..................................................................................... 3
1.1 Description conditions
3
1.2 List of safety instructions
3
2
System Description ................................................................................... 4
2.1 Functional description
4
2.2 Block Diagram
4
3
Quick Installation Guide ................................................
Operation Manual EMGZ 309
07.08.2017
3
1 Safety Instructions
1.1 Description conditions
a) High danger of health injury or loss of life
Danger
This symbol refers to high risk for persons to get health injury or loss of life. It has
to be followed strictly.
b) Risk of damage of machines
Caution
This symbol refers to informations, that, if ignored, could cause heavy mecanical
damage. This warning has to be followed absolutely.
c) Note for proper function
Not
Operation Manual EMGZ309 07.08.2017
4
2 System Description
Fig. 1: System Configuration EMGZ309 Tension Measuring Amplifier
E309012e
2.1 Functional description
The digital tension measuring amplifiers EMGZ309 are suitable for tension
measurement with all FMS force sensors. This versatile product comes as a plug &
play unit and is very easy to setup through the operating panel. Alternatively, more
advanced users can access all the internal parameters throug
Operation Manual EMGZ 309 07.08.2017
5
3 Quick Installation Guide
In a Plug & Play configuration the set-up of the EMGZ309 and force sensor is
limited to only the offset compensation and the calibration.
3.1 Preparations for Set-up
1. Read the Operation Manual of your force measuring sensors
2. Check your requirements such as:
– Desired output configuration 0…10V or 0/4… 20mA
– Display unit of nominal force
3. Draw the wiring diagram for your conf
Operation Manual EMGZ309 07.08.2017
6
Caution
Local installation regulations are to preserve safety of electric equipment. They
are not taken into consideration by this operating manual. However, they have to
be followed strictly.
3.4 Mounting the Force Sensors
Mounting of the force sensors is done according to the FMS Installation manual
which is delivered together with the force sensors.
3.5 Wiring the Amplifier
One or two force sensors can be connected
Operation Manual EMGZ 309 07.08.2017
7
Fig. 3: Wiring Diagram EMGZ309
EMGZ_309_BA_Manual.ai
Ethernet connectors
EMGZ_309_BA_Manual.ai
Caution
Bad earth ground connection may cause electric shock to persons, malfunction of
the total system or damage of the electronic unit! It is vital to ensure that proper
earth connection is done.
Note
The shield should be connected only to the electronic unit. On the force sensor
side the shield should stay open.
Operation Manual EMGZ309 07.08.2017
8
4 Configuring the Amplifier
4.1 Power up the EMGZ 309
1. Connect the first force sensor (see Fig.3 Wiring Diagram EMGZ309)
2. Check whether applying a force in measuring direction (in the direction of the
red point) on the first sensor results in a positive output signal. If not, exchange
the two signal wires of the corresponding force sensor in the terminal block
(terminals 12/13).
3. If applicable, connect the secon
Operation Manual EMGZ 309 07.08.2017
9
4.3 Change the Measuring Units
The default tension measuring units is N (Newton). The user can, however, select
another unit before the calibration process. To select the units of measure, follow
the steps below:
1. Hold the PARA key until the display changes to [Dis. Filt]
2. Scroll the menu by pressing the
key until the display shows [Unit]
3. Press the
and the word [Unit] will start flashing indicating that the
Operation Manual EMGZ309 07.08.2017
10
4.6 Offset Compensation over the Operating Panel
1. Press the two keys >0< and PARA at the same time for longer then 3 seconds.
2. The display will show [Offset Complete] when the process is finished (see Fig. 5
State diagram offset compensation).
Fig. 5: State diagram offset compensation
EMGZ309013e
4.7 Calibration
By calibrating (setting the Gain), amplifier and
load cells are adjust so that the display gives the
ac
Operation Manual EMGZ 309 07.08.2017
11
4.8 Calibration over the Operating Panel
1. Load a rope with a defined weight corresponding to your calibration force on
the roller. The roller configuration must correspond to the real configuration in
the machine (wrap angle, distance of the rollers etc.).
2. Enter calibration mode by pressing the two keys Cal and PARA at the same
time for longer than 3 seconds. The display changes to
3. The System Force [SysForce]
Operation Manual EMGZ309 07.08.2017
12
4.9 Analytic Calibration Method)
Determining the gain by means of the FMS-Calculator:
This method is used when the traditional calibration method with a weight is
difficult or impossible at all due to machine accessibility, weight adequateness etc.
This calculation method requires the FMS-calculator V2.1 or higher.
If necessary download and install FMS-calculator from the FMS homepage at:
http://www.fms-technology.com
Operation Manual EMGZ309 07.08.2017
14
5 Operation
5.1 Operating the EMGZ309 over the Operating Panel
The EMGZ309 amplifier can go in 5 different states:
• Operation State:
Amplifier operating
• Display Selection:
Selection of force or output reading
• Offset Compensation:
Automatic procedure to offset the roller weight
• Calibration State:
Procedure to calculate the amplifier gain
• Parameterisation State:
Set or change parameters
When using the
Operation Manual EMGZ 309 07.08.2017
15
5.2 Display Value Selection State
The EMGZ309 can display either a tension value or the amplifier output that is an
electrical unit corresponding to the measured tension.
A) For tension display following unit can be selected: N, kN, lb, g, kg
B) The amplifier output can be displayed in 0…10VDC, 0…20mA or 4…20mA
5.3 Displaying of Overload, Over and Underflow
In the Operation State the EMGZ309 amplifier will indi
Operation Manual EMGZ309 07.08.2017
16
6 Parameterization over the Operating Panel
After offset compensation and calibration the system is configured. It can now be
put into operation. The following state diagrams (Fig. 10 -12) help to navigate
through the parameter setting menu. In the Parameterisation State the EMGZ309
can be brought into the 2 modes: “Parameter Selection” and “Parameter Change”.
Fig. 10: State Diagram Parameter Setting (1st part
Operation Manual EMGZ 309 07.08.2017
17
1) The display shows the unit of measure that was previously selected.
3) Permanent key pressing expedites the changing speed.
4) If the Device Mode [VoltGauge] is set, these parameters are deactivated
Fig. 11: State Diagram Parameter Setting (2nd part)
E309017e