Microlife NC 400 Manual
Text of Microlife NC 400 Manual
IB NC 400 S-V11 3320
Revision Date: 2020-08-03
Microlife AG
Espenstrasse 139
9443 Widnau / Switzerland
www.microlife.com
NC400
Non Contact Thermometer
Microlife NC 400
1
2a
2b
3
4
EN
ES
FR
IT
DE
NL
GR
TR
AR
FA
2
10
18
26
34
52
60
70
78
86
PT
44
≤ 5cm
www.microlife.com
Name of Purchaser / Nom de l'acheteur / Nome del rivenditore / Name des Käufers
/ Alıcının adı / Nombre del comprador / Nome do comprador / Naam koper /
Ονοματεπώνυμο αγοραστή / /
Serial Number / Numéro de série / Numero di serie / Serien-Nr. / Seri numarası /
Número de serie / Número de série / Serienummer / Αριθμός σειράς /
/
Date of Purchase / Date d'achat / Data d’acquisto / Kauf
Table of Contents
1. The Advantages of this Thermometer
2. Important Safety Instructions
3. How this Thermometer measures Temperature
4. Control Displays and Symbols
5. Setting Date, Time and Beeper Functions
6. Changing between Body and Object Mode
7. Directions for Use
8. Changing between Celsius and Fahrenheit
9. How to recall 30 readings in Memory Mode
10.Error Messages
11.Cleaning and Disinfecting
12.Battery Replacement
13.Guarantee
14.Tec
A basic physiological effect called vasoconstriction can
occur in the early stages of fever, resulting in a cool skin
effect. The recorded temperature using this thermo-
meter can, therefore, be unusually low.
If the measurement result is not consistent with the patient's
finding or unusually low, repeat the measurement every 15
minutes or double check the result by another core body
temperature measurement.
This device comprises sensitive compo
3. Follow the previously mentioned instructions to set the
day, hours and minutes.
4. Once you have set the minutes and pressed the START/
IO button 7, the date and time are set and the time is
displayed.
If no button is pressed for 20 seconds, the device
automatically switches to ready for measuring AT.
Cancel time setup: Press the START/IO button
7 during time setup. The LCD will show Date/
Time icons with «--:--». After that press
6. For the next measurement wait until the «°C»/«°F» icon
is flashing and follow steps 4-5 above.
7. Press and hold the START/IO button 7 for 3 seconds
to turn off the device; otherwise the device will automat-
ically switch off after approx. 30 seconds.
NOTE:
Patients and thermometer should stay in similar
room condition for at least 30 minutes.
Don't take a measurement while or immediately after
nursing a baby.
Don't use the thermome
Pressing and releasing the MEMO-button 5 after the last
30 readings have been recalled will resume the above
sequence from reading 1.
10.Error Messages
Measured temperature too high AR: Displays «H»
when measured temperature is higher than 43 °C / 109.4
°F in body mode or 99.9 °C / 211.8 °F in object mode.
Measured temperature too low AS: Displays «L»
when measured temperature is lower than 34.0 °C /
93.2 °F in body mode or 0.1
returned with the original invoice. Repair or replacement
within guarantee does not prolong or renew the guarantee
period. The legal claims and rights of consumers are not
limited by this guarantee.
14.Technical Specifications
Type:
Non Contact Thermometer NC 400
Measurement
range:
Body mode: 34.0 - 43 °C / 93.2 -
109.4 °F
Object mode: 0.1 - 99.9 °C / 32.2 -
211.8 °F
Resolution:
0.1 °C / °F
Measurement
accuracy
(Laboratory):
Body mode:
±0.2 °C, 3
This device complies with the requirements of the Medical
Device Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial tech-
nical inspection is recommended for professional users.
Please observe the applicable disposal regulations.
15.www.microlife.com
Detailed user information about our thermometers and
blood pressure monitors as well as services can be found
at www.microlife.com.
Weight:
10