3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Table of Contents I. Important Safety Information ............................................ 3 II. Description of Contents ..................................................... 4 III. Set-Up Guide .................................................................... 5 IV. Quick Start Guide ............................................................. 8 V. Operating Instru

ETS 4406 D00939C - Page 3

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) I. Important Safety Information SAFETY INSTRUCTIONS The equipment described in this Manual is designed and manufactured to operate within defined design limits. Any misuse may result in electric shock or fire. To prevent the equipment from being damaged, the following rules should be observed for installation, use and maintenance. Read the following safety instructions befo

ETS 4406 D00939C - Page 4

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) II. Description of Contents Included in the standard Model 4406 package: Item Qty. Description Control Unit 1 Control unit houses the primary control and operating systems to provide high voltage, relay signals and sensor analysis. Faraday Cage 1 Cage provides the means to test various samples. Sensor (Model # 406- 104) 1 The Model 4406 includes one sensor as standard system

D00939C - Page 5

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) III. Set-Up Guide Diagram of Operating System Cables Step 1 – With AC power DISCONNECTED, Connect High Voltage Cable Make sure Power Switch on Control Unit’s front panel is in OFF position before inserting connectors. Connect the High Voltage Cable from the Faraday Cage to the Control Unit. Sensor HV Gnd Control Faraday Test Cage

D00939C - Page 6

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Step 2 – Connect Ground Cable Connect the black ground cable from the Faraday Cage to the Control Unit. Step 3 – Connect Sensor Cable Connect the grey sensor cable from the Faraday Cage to the Control Unit. Step 4 - Connect Test Unit Cable Connect the Test Unit Cable from the Faraday Cage to the Control Unit.

D00939C - Page 7

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Step 5 - Connect Power Cable Make sure Power Switch on Control Unit’s front panel is in OFF position. Connect the Power Cable to the back of the Control Unit. Connect the Control Unit to the AC line by plugging its AC cord into a properly grounded AC outlet (Mains). The Power Cable may be plugged into any compatible outlet receptacle with 90-260 volts.

D00939C - Page 8

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) IV. Quick Start Guide Step 1 – Switch Power ON Turn on POWER and allow the system to warm up for at least 5 minutes prior to testing. (If the Analyzer is to be used on an ongoing basis it can remain powered-up, but should be placed in the ZERO/STBY mode.) Step 2 – Open Faraday Cage cover and install STM-2 module as shown Use the pair of rod-shaped Magnets to hold the STM

D00939C - Page 9

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Step 4 – Press “SYSTEM TEST” to run balance tests Step 5 – Wait until “SYSTEM TEST” shows “PASSED” Step 6 – Open Faraday Cage cover and remove STM-2. Remove the pair of rod-shaped Magnets holding the STM-M module in place. Be careful to avoid allowing the STM-2 to fall when removing the 2nd Magnet to avoid damage.

D00939C - Page 10

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Step 7 – Install test sample, then close Faraday cage cover After removing the STM-2, use either the Magnets or Clamp Electrodes to secure the test sample. Step 8 – Press “MAN/AUTO” to go to AUTO mode Step 9 – Make sure display shows AUTO mode

D00939C - Page 11

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Step 10 – Press “DECAY TEST” to start AUTO tests Step 11 – Wait for “TEST OVER. Record data then repeat steps 7-10 for next sample, etc.

ETS 4406 D00939C - Page 12

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) V. Operating Instructions A. Initial Turn-On & Verification When turned on the system will default to ZERO/STBY status and the settings from any previous testing will be displayed as shown in Figure 5-1. Figure 5-1: Typical ZERO/STBY displays at initial turn-on At this time the Charging Voltage status, Test Mode and Cutoff level for sample testing can be set as follows: a.

D00939C - Page 13

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) d. Full Scale Calibration Prior to testing it is necessary to check (or adjust) the full scale calibration of the electrostatic measurement. This is normally performed using the STM-2 as follows: Lift the hinged cover of the Faraday Cage and place the STM-2 Module between the electrodes of the sample holder as shown in Figure 5-2. Return the cover to the operating position

D00939C - Page 14

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) NOTE: The unit employs a safety interlock system which prevents high voltages from being impressed on the test sample unless the Faraday Cage cover is positioned in the operate (cover down) position. Voltages up ±5000 Volts (current is limited to 50 microamps) are generated by the Control Unit and are transferred to the electrodes and subsequently to the test sample during

D00939C - Page 15

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) If the system is not balanced, it will automatically try to rebalance and the display will read as shown in Figure 5-4. Figure 5-4: Out of balance Auto Adjust If the system does not auto balance and/or the timer runs on, the display will indicate SYSTEM TEST Aborted as shown in Figure 5-5. Figure 5-5: System fails Self Test If this message is displayed, the Analyzer may nee

D00939C - Page 16

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) When testing a sample with unknown static dissipation characteristics it is necessary that its basic characteristics be measured by observing the existence of an initial charge (IC) and its ability to conduct on the charge applied to the electrodes. This analysis is performed by depressing the CHARGE key. Any charge detected on the surface of the sample will register as IC=

D00939C - Page 17

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Figure 6-2: Display for a sample that won’t charge If the sample has some limited static dissipative characteristics the accepted charge after applying +5.00V for 60 seconds is displayed. The more static dissipative the material the greater the amount of charge accepted. This measurement is very helpful when trying to evaluate materials with antistat additive levels that

D00939C - Page 18

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) 3.74 kV. The second display is for a sample that also exhibits static dissipative characteristic, but was a full size sample. Figure 6-4: Sample with no IC and able to conduct charge d. A composite or laminated material will exhibit similar characteristics, but the apparent accepted charge will be more like 3.50-4.75kV. If a decay time measurement is attempted the time wi

ETS 4406 D00939C - Page 19

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) Table 6.1 Cutoff Levels and Applicable Specifications *Nominal value. Actual cutoff is the percentage selected by the CUTOFF level selected. Example: If accepted voltage is 4.5kV, cutoff is 450V. e. Decay Test MANual mode: Depress the DECAY TEST key. The electrostatic sensor shutter will close momentarily (an audible click should be heard) then open to start the decay ti

D00939C - Page 20

3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878) To repeat or test a new sample, go to ZERO/STBY or depress the CHARGE or DECAY TEST key. AUTO mode: Select the AUTO mode by depressing the AUTO/MAN key. Depress the DECAY TEST key to start an automatic static decay time measurement sequence. Up to 6 measurements at +V and then at –V can be programmed. Three (3) measurements at each polarity is the default setting. Both +V

ETS Measuring Instruments Owners Instructions: