ETS 4406 Operating Manual
Text of ETS 4406 Operating Manual
D00939C - Page 2
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Table of Contents
I. Important Safety Information ............................................ 3
II. Description of Contents ..................................................... 4
III. Set-Up Guide .................................................................... 5
IV. Quick Start Guide ............................................................. 8
V. Operating Instru
ETS 4406 D00939C - Page 3
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
I. Important Safety Information
SAFETY INSTRUCTIONS
The equipment described in this Manual is designed and manufactured to operate within defined
design limits. Any misuse may result in electric shock or fire. To prevent the equipment from being
damaged, the following rules should be observed for installation, use and maintenance. Read the
following safety instructions befo
ETS 4406 D00939C - Page 4
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
II. Description of Contents
Included in the standard Model 4406 package:
Item
Qty.
Description
Control Unit
1
Control unit houses the primary control and operating systems to
provide high voltage, relay signals and sensor analysis.
Faraday Cage
1
Cage provides the means to test various samples.
Sensor (Model # 406-
104)
1
The Model 4406 includes one sensor as standard system
D00939C - Page 5
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
III. Set-Up Guide
Diagram of Operating System Cables
Step 1 – With AC power
DISCONNECTED, Connect High
Voltage Cable
Make sure Power Switch on Control
Unit’s front panel is in OFF position
before inserting connectors.
Connect the High Voltage Cable from the
Faraday Cage to the Control Unit.
Sensor
HV Gnd
Control
Faraday Test Cage
D00939C - Page 6
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Step 2 – Connect Ground Cable
Connect the black ground cable from the
Faraday Cage to the Control Unit.
Step 3 – Connect Sensor Cable
Connect the grey sensor cable from the
Faraday Cage to the Control Unit.
Step 4 - Connect Test Unit Cable
Connect the Test Unit Cable from the
Faraday Cage to the Control Unit.
D00939C - Page 7
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Step 5 - Connect Power Cable
Make sure Power Switch on Control
Unit’s front panel is in OFF position.
Connect the Power Cable to the back of
the Control Unit.
Connect the Control Unit to the AC line
by plugging its AC cord into a properly
grounded AC outlet (Mains). The Power
Cable may be plugged into any
compatible outlet receptacle with 90-260
volts.
D00939C - Page 8
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
IV. Quick Start Guide
Step 1 – Switch Power ON
Turn on POWER and allow the system to warm
up for at least 5 minutes prior to testing. (If the
Analyzer is to be used on an ongoing basis it can
remain powered-up, but should be placed in the
ZERO/STBY mode.)
Step 2 – Open Faraday Cage cover and install
STM-2 module as shown
Use the pair of rod-shaped Magnets to hold the
STM
D00939C - Page 9
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Step 4 – Press “SYSTEM TEST” to run balance
tests
Step 5 – Wait until “SYSTEM TEST” shows
“PASSED”
Step 6 – Open Faraday Cage cover and
remove STM-2.
Remove the pair of rod-shaped Magnets holding
the STM-M module in place. Be careful to avoid
allowing the STM-2 to fall when removing the 2nd
Magnet to avoid damage.
D00939C - Page 10
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Step 7 – Install test sample, then
close Faraday cage cover
After removing the STM-2, use either the Magnets
or Clamp Electrodes to secure the test sample.
Step 8 – Press “MAN/AUTO” to go to
AUTO mode
Step 9 – Make sure display shows AUTO mode
D00939C - Page 11
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Step 10 – Press “DECAY TEST” to start
AUTO tests
Step 11 – Wait for “TEST OVER. Record data
then repeat steps 7-10 for next sample, etc.
ETS 4406 D00939C - Page 12
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
V. Operating Instructions
A. Initial Turn-On & Verification
When turned on the system will default to ZERO/STBY status and the settings from any previous
testing will be displayed as shown in Figure 5-1.
Figure 5-1: Typical ZERO/STBY displays at initial turn-on
At this time the Charging Voltage status, Test Mode and Cutoff level for sample testing can be set
as follows:
a.
D00939C - Page 13
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
d. Full Scale Calibration
Prior to testing it is necessary to check (or adjust) the full scale calibration of the electrostatic
measurement. This is normally performed using the STM-2 as follows:
Lift the hinged cover of the Faraday Cage and place the STM-2 Module between the electrodes of
the sample holder as shown in Figure 5-2. Return the cover to the operating position
D00939C - Page 14
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
NOTE: The unit employs a safety interlock system which prevents high voltages from being
impressed on the test sample unless the Faraday Cage cover is positioned in the operate (cover
down) position. Voltages up ±5000 Volts (current is limited to 50 microamps) are generated by the
Control Unit and are transferred to the electrodes and subsequently to the test sample during
D00939C - Page 15
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
If the system is not balanced, it will automatically try to rebalance and the display will read as
shown in Figure 5-4.
Figure 5-4: Out of balance Auto Adjust
If the system does not auto balance and/or the timer runs on, the display will indicate SYSTEM
TEST Aborted as shown in Figure 5-5.
Figure 5-5: System fails Self Test
If this message is displayed, the Analyzer may nee
D00939C - Page 16
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
When testing a sample with unknown static dissipation characteristics it is necessary that its
basic characteristics be measured by observing the existence of an initial charge (IC) and its
ability to conduct on the charge applied to the electrodes. This analysis is performed by
depressing the CHARGE key. Any charge detected on the surface of the sample will register as
IC=
D00939C - Page 17
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Figure 6-2: Display for a sample that won’t charge
If the sample has some limited static dissipative characteristics the accepted charge after
applying +5.00V for 60 seconds is displayed. The more static dissipative the material the
greater the amount of charge accepted. This measurement is very helpful when trying to
evaluate materials with antistat additive levels that
D00939C - Page 18
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
3.74 kV. The second display is for a sample that also exhibits static dissipative characteristic, but
was a full size sample.
Figure 6-4: Sample with no IC and able to conduct charge
d. A composite or laminated material will exhibit similar characteristics, but the apparent accepted
charge will be more like 3.50-4.75kV. If a decay time measurement is attempted the time wi
ETS 4406 D00939C - Page 19
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
Table 6.1 Cutoff Levels and Applicable Specifications
*Nominal value. Actual cutoff is the percentage selected by the CUTOFF level selected.
Example: If accepted voltage is 4.5kV, cutoff is 450V.
e. Decay Test
MANual mode: Depress the DECAY TEST key. The electrostatic sensor shutter will close
momentarily (an audible click should be heard) then open to start the decay ti
D00939C - Page 20
3101 Mt Carmel Ave. Glenside, PA │ 833-ESD-GURU (833-373-4878)
To repeat or test a new sample, go to ZERO/STBY or depress the CHARGE or DECAY TEST
key.
AUTO mode: Select the AUTO mode by depressing the AUTO/MAN key. Depress the DECAY
TEST key to start an automatic static decay time measurement sequence. Up to 6 measurements
at +V and then at –V can be programmed. Three (3) measurements at each polarity is the default
setting. Both +V