ETS 212 Operating Manual
Text of ETS 212 Operating Manual
ETS 212 Electro-Tech Systems, 700 West Park Avenue, Perkasie, PA 19844 - 833-ENV-GURU (833-368-4878)
D00970 Revision C
ETS STATIC METER - Model 212
SURVEYORSTAT - Model 212XL
Operating Manual
Model 212 – Static Meter – Operating Manual
D00970 Revision C - Page 2 of 15
Contents
I. Important Safety Information .............................................................................................. 3
II. Description of Components ............................................................................................... 4
III. Set-Up Guide .............................................................................................
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 3 of 15
I. Important Safety Information
SAFETY INSTRUCTIONS
The equipment described in this Manual is designed and manufactured to operate within
defined design limits. Any misuse may result in electric shock or fire. To prevent the equipment
from being damaged, the following rules should be observed for installation, use and
maintenance. Read the following safety instructions befor
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 4 of 15
II. Description of Components
Included:
Item
Qty.
Description
212XL or 212
1
Static meter (212XL includes separate sensor and extension arm)
Cable
1
Black 10’ coiled cord with clip
Battery
1
(Installed)
Case
1
(212 only – Black vinyl case)
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 5 of 15
III. Set-Up Guide
Inspection and Power Up.
□ Tools Needed: None
□ Make sure all power is OFF.
Step 1 – Inspect Equipment
For the 212XL, be sure all cables are
securely fastened..
Step 2 – Power ON
Slide the power switch to the ON position and
verify activity on the display.
Step 3 – Battery check
With power ON, check the display. Battery
low is indicated by all 4
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 6 of 15
IV. Quick Start Guide
□
Tools Needed: None
Step 1 – Turn on Power
Turn on Power. Take note of display
showing 0.00 and no battery low indication.
Step 2 – Ground the 212.
Attach the Ground Cable to the snap on the
back of the case. Connect the other end to a
known ground. If holding the 212 in your
bare hand, discharge by touching ground.
Step 3 – Set 1 inch dista
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 7 of 15
Step 5 – HOLD last reading
With the Meter positioned 1 inch from the object
being measured, press the MEASURE/HOLD
pushbutton so that it releases to the upper or
HOLD position. This will freeze the reading and
allows the operator to move the Meter where it
may be more easily read, or saved for later
reference. In the HOLD position the red LED will
light. The analog output
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 8 of 15
V. Operating Instructions
The 212 Product Family
The Model 212 and 212XL Static Meters are accurate, compact electrostatic field meters used
for locating and measuring static charge potentials.
The Model 212 is a handy, compact handheld unit which utilizes a 3½-digit LCD digital display
readout. Ranging lights assure accurate and repeatable measurements when measuring flat
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 9 of 15
MAKING A MEASUREMENT
Hold the Meter 1 inch (25mm) from the object to be measured. This distance is measured
from the front edge of the Meter case to the surface of the object. The Meter will display a
reading of the electrostatic field in kilovolts per inch from minus 19.99kV to plus 19.99kV.
In the MEASURE position, the RANGING LIGHTS are on. These lights help place the Met
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 10 of 15
MEASUREMENTS USING THE 212XL
The 212XL Surveyorstat is designed specifically to measure static charges on moving webs
or hard to reach locations.
CAUTION: When taking measurements around moving machinery extreme care must
be taken to ensure the sensor probe does not interfere or get caught in the process.
The sensor used in these meters is very fragile and sensitive to mech
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 11 of 15
Front Panel Description and function
Device
Description
Functionality
1
ON-OFF Switch
This switch disconnects the battery.
2
MEASURE Button
This is a “click-ON/click OFF” type of pushbutton
3
ZERO adjustment
Rotary knob used for fine adjustment of the 0.00 level.
4
Display
3-1/2 digit LCD digital display.
5
REC connector
2.5mm phone jack for analog recorder output.
6
R
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 12 of 15
SPECIFICATIONS
Model 212
Display
LCD. 3½-digit numeric display, 0.375” digit height. Displays measured voltage
in kV when measurement distance is 1”.
Range
-19.99 kV to +19.99 kV at a distance of 1 inch (25mm). Higher voltages may be
measured at distances greater than 1 inch
Accuracy:
±5%
Controls:
ON/OFF slide switch. MEASURE/HOLD pushbutton (latching push-ON/push-
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 13 of 15
VI. Calibration and Maintenance
CALIBRATION is not required. The Meters are factory calibrated.
MAINTENANCE
No routine maintenance is required. If for any reason the Meter is not working correctly, contact ETS at
215-887-2196 ext. 220 for assistance. NOTE: There are no user serviceable parts. Any unauthorized
service will void the warranty and result in additional repair
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 14 of 15
Place the Static Meter on the support and locate it exactly 1” (25.4mm) from the metal plate. Turn
on the Static Meter and adjust the ZERO if necessary. Turn on the calibrated power supply. The
Meter should read the applied voltage 5% for the Model 212. For the Model 211, the correct
indicator should light. Small adjustments of the Meter distance from the plate will ch
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 15 of 15
APPENDIX A – Probes and Detectors
Model 205C, 205C-x1 and 205C-x10 Charged Plate Detectors
These charged plate detectors attach to the front of the Model 211 and 212 Static Meter. The
standard Model 205C along with the optional 6”x6” Detector Plate is shown in below increases
the sensitivity of the measurement system by a factor of 10 and is designed for low voltage
m