ETS 212 Operating Manual

Text of ETS 212 Operating Manual

ETS 212 Electro-Tech Systems, 700 West Park Avenue, Perkasie, PA 19844 - 833-ENV-GURU (833-368-4878)

D00970 Revision C ETS STATIC METER - Model 212 SURVEYORSTAT - Model 212XL Operating Manual

Model 212 – Static Meter – Operating Manual

D00970 Revision C - Page 2 of 15 Contents I. Important Safety Information .............................................................................................. 3 II. Description of Components ............................................................................................... 4 III. Set-Up Guide .............................................................................................
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 3 of 15 I. Important Safety Information SAFETY INSTRUCTIONS The equipment described in this Manual is designed and manufactured to operate within defined design limits. Any misuse may result in electric shock or fire. To prevent the equipment from being damaged, the following rules should be observed for installation, use and maintenance. Read the following safety instructions befor
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 4 of 15 II. Description of Components Included: Item Qty. Description 212XL or 212 1 Static meter (212XL includes separate sensor and extension arm) Cable 1 Black 10’ coiled cord with clip Battery 1 (Installed) Case 1 (212 only – Black vinyl case)
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 5 of 15 III. Set-Up Guide Inspection and Power Up. □ Tools Needed: None □ Make sure all power is OFF. Step 1 – Inspect Equipment For the 212XL, be sure all cables are securely fastened.. Step 2 – Power ON Slide the power switch to the ON position and verify activity on the display. Step 3 – Battery check With power ON, check the display. Battery low is indicated by all 4
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 6 of 15 IV. Quick Start Guide □ Tools Needed: None Step 1 – Turn on Power Turn on Power. Take note of display showing 0.00 and no battery low indication. Step 2 – Ground the 212. Attach the Ground Cable to the snap on the back of the case. Connect the other end to a known ground. If holding the 212 in your bare hand, discharge by touching ground. Step 3 – Set 1 inch dista
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 7 of 15 Step 5 – HOLD last reading With the Meter positioned 1 inch from the object being measured, press the MEASURE/HOLD pushbutton so that it releases to the upper or HOLD position. This will freeze the reading and allows the operator to move the Meter where it may be more easily read, or saved for later reference. In the HOLD position the red LED will light. The analog output
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 8 of 15 V. Operating Instructions The 212 Product Family The Model 212 and 212XL Static Meters are accurate, compact electrostatic field meters used for locating and measuring static charge potentials. The Model 212 is a handy, compact handheld unit which utilizes a 3½-digit LCD digital display readout. Ranging lights assure accurate and repeatable measurements when measuring flat
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 9 of 15 MAKING A MEASUREMENT Hold the Meter 1 inch (25mm) from the object to be measured. This distance is measured from the front edge of the Meter case to the surface of the object. The Meter will display a reading of the electrostatic field in kilovolts per inch from minus 19.99kV to plus 19.99kV. In the MEASURE position, the RANGING LIGHTS are on. These lights help place the Met
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 10 of 15 MEASUREMENTS USING THE 212XL The 212XL Surveyorstat is designed specifically to measure static charges on moving webs or hard to reach locations. CAUTION: When taking measurements around moving machinery extreme care must be taken to ensure the sensor probe does not interfere or get caught in the process. The sensor used in these meters is very fragile and sensitive to mech
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 11 of 15 Front Panel Description and function Device Description Functionality 1 ON-OFF Switch This switch disconnects the battery. 2 MEASURE Button This is a “click-ON/click OFF” type of pushbutton 3 ZERO adjustment Rotary knob used for fine adjustment of the 0.00 level. 4 Display 3-1/2 digit LCD digital display. 5 REC connector 2.5mm phone jack for analog recorder output. 6 R
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 12 of 15 SPECIFICATIONS Model 212 Display LCD. 3½-digit numeric display, 0.375” digit height. Displays measured voltage in kV when measurement distance is 1”. Range -19.99 kV to +19.99 kV at a distance of 1 inch (25mm). Higher voltages may be measured at distances greater than 1 inch Accuracy: ±5% Controls: ON/OFF slide switch. MEASURE/HOLD pushbutton (latching push-ON/push-
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 13 of 15 VI. Calibration and Maintenance CALIBRATION is not required. The Meters are factory calibrated. MAINTENANCE No routine maintenance is required. If for any reason the Meter is not working correctly, contact ETS at 215-887-2196 ext. 220 for assistance. NOTE: There are no user serviceable parts. Any unauthorized service will void the warranty and result in additional repair
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 14 of 15 Place the Static Meter on the support and locate it exactly 1” (25.4mm) from the metal plate. Turn on the Static Meter and adjust the ZERO if necessary. Turn on the calibrated power supply. The Meter should read the applied voltage 5% for the Model 212. For the Model 211, the correct indicator should light. Small adjustments of the Meter distance from the plate will ch
Model 212 – Static Meter – Operating Manual D00970 Revision C - Page 15 of 15 APPENDIX A – Probes and Detectors Model 205C, 205C-x1 and 205C-x10 Charged Plate Detectors These charged plate detectors attach to the front of the Model 211 and 212 Static Meter. The standard Model 205C along with the optional 6”x6” Detector Plate is shown in below increases the sensitivity of the measurement system by a factor of 10 and is designed for low voltage m

ETS Measuring Instruments Owners Instructions: