Radiodetection RD500 Operation Manual
Text of Radiodetection RD500 Operation Manual
2 | P a g e
Before you start
Thank you for your interest in Radiodetection’s
RD500 Plastic Pipe Locator Kit
Please read this user manual in its entirety
before attempting to use the RD500 system as
it contains many important safety notices and
warnings.
Radiodetection products, including this manual,
are under continuous development. The
information contained within is accurate at the
time of publication; however the RD500, this
manual and all its conten
3 | P a g e
Section
Title
Page
INTRODUCTION
PART 1 - OPERATING GUIDE
5
1.0
Unpacking and Setup
5
2.0
Fast Start Guide
5
3.0
Pulse Transmitter Application Techniques
6
4.0
Locator Application Techniques
7
5.0
Locator & Adapter Kit Details
8
5.1
Locator Kit Details
8
5.2
Sprinkler/Hose Bib Kit Details
9
5.3
Fire Hydrant Kit Details
9
5.4
RD500-Tx Electronic Pulse Transmitter
10
5.5
Water Meter Kit Details
10
PART II - TROUBLESHOOTING GUIDE
1.0
Introduc
4 | P a g e
List of Figures & Tables
Figure
Title
Page
PART 1 - FIGURES
5.1
Locator Kit Details
8
5.2
Sprinkler/ Hose Bib Kit Details
9
5.3
Fire Hydrant Kit Details
9
5.4
RD500-TX kit details
10
5.5
Water Meter Standpipe Kit Details
10
PART II - FIGURES
2.1
Pulse Transmitter Photograph
11
2.2
Pulse Transmitter Simplified Block Diagram
12
3.1
Locator Photograph
17
3.2
RD500 Theory and use
21
LIST OF TABLES
Table 2.1
Pulse Transmitter Type S
5 | P a g e
INTRODUCTION
Thank you for purchasing the most cost-effective plastic pipe locator tool available. The purpose of this
manual is to provide useful information that will help you get the best possible results from your new
locator. This operating guide contains directions for unpacking, setting up and operating the RD500
system. It also contains a fast start guide that will enable users to quickly begin locating water pipes. This
is followed by
6 | P a g e
4.
Begin initial tracing about 2 metres from Pulse Transmitter or any other area where
pipe location is known
5.
Push spike of Locator into ground near known pipe location and turn ON/ OFF knob
clockwise for mid-range meter reading and adjust individual headphone controls for
comfortable listening level
6.
To pinpoint pipe position, use Locator to probe ground until you find peak
(maximum) meter reading. You will also hear the pulsing sound (si
7 | P a g e
Check the water pressure before attempting to locate the water line and make sure that it is in the
operating range of the Pulse Transmitter, which is 3 to 10 Bars (40 to 140 psi).
The Pulse Transmitters are color-coded to identify their intended usage: A color-coded ty-wrap is used
along with a small square area that is painted on the side of the Pulse Transmitter to indicate the color.
Green is the lowest power Pulse Transmitter and is intend
Radiodetection RD500 8 | P a g e
Best location results are obtained by beginning your tracing at least 3 metres from the location where
the Pulse Transmitter is connected. The signal levels are normally too strong in the area immediately
surrounding the Pulse Transmitter. Alternately, tracing may begin at a known pipe location that is not
more than 75 metres down the pipe from the Pulse Transmitter.
The preferred method for tracing pipes is to locate at least two known pipe lo
11 | P a g e
Lock Nut
Serial Number
Outlet
Coloured Ty-wrap
Inlet
PART II - TROUBLESHOOTING GUIDE
1.0
Introduction:
This part of the manual is divided into two major subjects. Section 2 covers the Pulse Transmitter and
Section 3 covers the Locator. Each section is organized to first provide an overview of the equipment
and its major components. This is followed by a set of tests that can be performed by the user to locate
problems.
2.0
Mechanical Pulse Tra
12 | P a g e
Figure 2.2 Pulse Transmitter Simplified Block Diagram
Figure 2.2 is a simplified block diagram of the Pulse Transmitter illustrating the operation of the valve.
Valve opening and closing is controlled by the adjustable T-handle. Turning the handle clockwise
increases the spring force and causes the valve to open (water starts flowing), while rotating it
anticlockwise reduces the spring force allowing it to close (water stops flowing). A handle
13 | P a g e
Larger diameter pipes require higher flow rates to produce a strong signal output. The three types of
Pulse Transmitters currently offered are summarized in Table 2.1 below:
Table 2.1 - Pulse Transmitter Type Summary
Power/ Color Code
Diameter of Pipe
Service Type
Low / GREEN
Medium / YELLOW
High / RED*
≤ 5cm
> 5cm
> 15cm
Service and Sprinkler lines
Small Water Main
Large Water Mains
*CAUTION: RED PULSE TRANSMITTERS SHOULD ONLY BE USED ON
14 | P a g e
2.2.1.2 Leak Tests:
1. Connect the Pulse Transmitter to the appropriate water pipe using all of the adapter kit
provided. DO NOT connect the water hose to the Pulse Transmitter outlet during this test.
2. Turn the water fully on and turn the T-handle on the Pulse Transmitter anticlockwise until
water stops flowing completely.
3. Observe the area around the inlet of the Pulse Transmitter. If any leaks occur, check the inlet
filter screen washer
15 | P a g e
2. Turn the water fully on (open valve completely).
3. Loosen the T-handle collar (knurled ring or hex-shaped collar just above T-handle). Turn
the T-handle on the Pulse Transmitter clockwise until water starts flowing through the water
hose on the outlet.
4. Slowly turn the T-handle anticlockwise until the Pulse Transmitter starts to pulse.
5. Turn the T-handle clockwise until the pulsing stops and observe its position (rotational
location) a
16 | P a g e
Table 2.3 RD500-Tx. Electronic Pulse transmitter failure symptoms and causes
SYMPTOM OBSERVED
LIKELY CAUSE (S)
No pulse from electronic Pulse transmitter
when Power is switched on
Battery flat , Check indicator on RD500-Tx
power Control unit
Check 12v DC supply is connected
3.0 RD500 Locator Repairs
3.1
Purpose of the RD500 Locator:
The RD500 Locator picks up the signal created inside the pipe by the Pulse Transmitter. The
Pulse Transmi
Radiodetection RD500 17 | P a g e
seal the upper control housing and a decorative black bezel is installed over the face of the
meter. The meter assembly is connected to the control housing by a single 8-pin connector.
3.2.1 RD500 Locator Tests:
The following tests can be easily performed by a user to help determine if a problem(s) exists
in the RD500 Locator. Please see Figure 3.2 to clarify the location of specific parts.
3.2.1.1 Battery Tests:
The RD500 Locator battery may
Radiodetection RD500 18 | P a g e
3.2.1.2 Locator Sensitivity Tests:
Each RD500 Locator Assembly is calibrated at the factory to produce specific meter signal levels when
the complete assembly is tested with a platform that produces controlled levels of motion. The following
test can be conducted without special equipment to determine if your Locator has sensitivity problems
and should be returned to Radiodetection for repair.
NOTE: the ON/ OFF SENSITIVITY Switch (located dire
19 | P a g e
9. If the meter does not return below 10% quickly after being tapped, and you hear a tone, then
the Locator is breaking into oscillation and may need to be returned to Radiodetection for
repair.
10.
If a tone (oscillation) is heard, try reducing the sensitivity by turning the knob
anticlockwise about 1/ 4 turn. If this does not help, watch the meter and observe its response
when oscillating, then try disconnecting the headphones and repeating