ETS 4431-EV Operating Manual
Text of ETS 4431-EV Operating Manual
SHIELDED BAG TEST SYSTEM
Model 4431-EV (Voltage & Energy)
&
Model 4431-V (Voltage Only)
(These models replace Model 431)
Operating Manual
10/10
(Equipment containing HV)
The equipment described in this Manual is designed and manufactured to operate within
defined design limits. Any misuse may result in electric shock or fire. To prevent the
equipment from being damaged, the following rules should be observed for installation, use
and maintenance. Read the following safety instructions before operating the instrument.
Retain these instructions in a safe place for fut
DO NOT OPERATE IN WET/DAMP CONDITIONS: If water or other liquid penetrates the
equipment, unplug the power cord and contact the factory for further instructions.
Continuous use in this case may result in electrical shock, fire or permanent damage to the
equipment.
DO NOT OPERATE IN HIGH HUMIDITY: Operating the equipment in high humidity
conditions will cause deteriation in performance, system failure, or present a shock or fire
hazard. Contact the factor
The rapid advancement in the electronics industry during the past decade has placed
increasing importance on the understanding of electrostatics and its effect on electronic
devices and systems. Electrostatic Discharge (ESD) is a common cause of
microelectronic circuit failure. Many of these devices can be seriously damaged or
destroyed by an electrostatic discharge below 30 Volts, or as a result of an electrostatic
field of only a few h
2.0 TEST METHODS
2.1
ANSI/ESD
STM11.31-2006
ESD
Association
standard
for
evaluating the performance of electrostatic discharge shielding
materials - Bags
This test method evaluates the performance of electrostatic discharge shielding
bags. The purpose of this standard is to ensure that testing laboratories, using
this test method to evaluate a given packaging material, will obtain similar
results. This standard specifies the discharge waveform characteri
Figure 2.0-3: Typical Model 4431 shielded bag waveform
Six specimens of a given sample are required. Six measurements per specimen
is specified. Testing is to be performed at both 12% and 50% ±3% RH at a
temperature of 73 ±3°F after a conditioning period of 48 hours minimum. Bag
size for this test should be 8x10 inches (20x25 cm).
ANSI/ESD STM11.31 specifies that the test conditions, peak current, minimum,
maximum and average energy levels of all 36 m
Figure 2.0-4: Voltage waveform for non-shielding and shielded bags
A good static shielding bag typically will have a voltage differential of 5 to 30 Volts.
Bags with voltage differentials above 150 Volts usually have a surface resistance
above the specified 1x103 Ohms upper limit. The voltage differential created by a
static shielding bag is primarily caused by the time delay for the applied pulse to be
induced on the lower electrode created by the resis
ETS 4431-EV Figure 3.0-1b: Model 4431-EV Shielded Bag Test System, laptop computer
This is a turn-key system that performs both the ANSI/ESD STM11.31 and the
EIA 541-1998 shielding bag tests. Depending on the test method selected the
System provides the correct discharge current pulse for both the ANSI/ESD
STM11.31 and EIA 541 tests. The capacitive sensor detects the current pulse or
the voltage differential and sends the signal(s) to the included oscilloscope for
d
ETS 4431-EV 9. Adjustable bag insertion stops for configuring the System for different
bag sizes per specification.
10. Calibration jumpers and cables
1.1.2 Oscilloscope
1. Tektronix Model TDS2022B
3.1.2 Computer (Desktop or Laptop)
1. Running Windows� 2000/XP/VISTA/7
3.1.3 Software
1. ETS Test Suit Manager, Energy Calculation Program (Version 4.0.3.3)
3.2. Model 4431-V
The Model 4431-V Discharge Unit is identical to the Model 4431-EV except it
does not include th
a foot operated switch, which is convenient when testing in a humidity chamber. In the
AUTO mode, the discharge sequence specified in ANSI/ESD STM11.31 along with the
scope settings are controlled automatically by the ETS Energy Calculation Program.
The discharge waveform generated by the Model 4431 meets the requirements of
ANSI/ESD STM5.1 Electrostatic Discharge Testing - Human Body Model. A typical
discharge waveform measured with the discharge electr
5. USB to serial converter with 9-pin sub-D from computer to communications
cable (Model 4431-EV only) NOTE: Do not sustitute
6. 9-pin sub-D from Converter to 25-pin sub-D to Model 4431-EV
7. Foot switch
8. IEC Line Cord (North American plug)
Figure 4.0-3b shows the jumpers and cables required to perform first time set up.
Figure 4.0-3b: Calibration cables and jumpers for first time set up
NOTE: If the Model 4431 is used to replace a Model 431 discharge
Figure 5.0-1: System set up for Energy test
1. Connect the BNC connector from the built-in CT-1 current transducer into
CH-1 of the oscilloscope. The Terminator provided must be used in order
to match the 50 Ohm output impedance of the CT-1 to the 1 MegOhm
impedance of the scope.
2. Connect the 25 pin – 9-pin cable to the USB – 9-pin adapter and 25-pin sub-
D connector on the rear of the Model 4431-EV
3. Connect the USB - USB cable to the 1st comput
Figure 5.0-2: Voltage differential test set up
If the Model 4431-EV is used to measure voltage differential, the CT-1 current
sensor must be disconnected and the Voltage cables connected to perform this
test. The Model 4431-EV is shipped from the factory with the capacitive sensor
programmed for the Energy test. To convert the System from energy to voltage
differential, the capacitive sensor must first be converted by plugging the red
jumpers into the c
ETS 4431-EV 1 Megohm (10,000 and 10 MegOhms/sq). Current industry practice is to
use 1500 Ohms.
5.2.2 Oscilloscope
The oscilloscope and Discharge Unit must first be calibrated as a set. If
the oscilloscope is supplied as part of the system, the Discharge Unit and
scope will have already been calibrated. If the scope is provided by the
user, it is first necessary to compensate the built in 100:1 attenuators and
cables to the scope input using the following procedure
Figure 5.0-4: Measured waveforms
Repeat the above for CH-2.
The amplitude of each signal must be the same. If not, adjust the CH-1
and/or 2 “R” controls to obtain equal signals.
After calibrating each channel individually, invert the CH-2 scope input
and select ADD. The composite waveform should look like that shown in
Figure 5.0-5c. Slight “R” and “C” adjustments may be required t
ETS 4431-EV especially if a digital storage oscilloscope is used, causing significant
measurement error. This condition normally occurs when buried metal
layer bags are tested and 5, 10 or 20 Volt sensitivity levels are selected to
obtain greater voltage differential resolution. To verify the signals, switch
out of the ADD mode and into the CHOP mode and observe the individual
CH-1 and CH-2 signals.
d. Set the trigger to NORM and CH-1 trigger source. Adjust the lev
18 gage wire
Figure 6.0-1: Calibration set up to measure ESD pulse
6.1.2 EIA 541-1988 Voltage Differential
EIA 541 does not specify the discharge waveform characteristics.
However, it is implied that the waveform meets the requirements of
ANSI/ESD STM 5.1. To verify the discharge waveform of the Model 4431-
V, follow the procedure described in paragraph 4 in Section 5.1.1 above.
6.2
Product Testing per ANSI/ESD STM11.31
ANSI/ESD STM11.31
Set the oscilloscope vertical sensitivity to 50mV/div, time base to 50nsec/div and
trigger level to 50mV. Trigger mode is NORMAL.
Turn on the HV.
Initiate a discharge, using either the front panel TEST switch or the foot switch.
The discharge pulse will activate for approximately 250 milliseconds. The current
pulse detected by the capacitive sensor will be displayed on the oscilloscope.
Typical waveforms for a static shielding bag are shown in Figure 6.
Figure: 6.0-4: Energy Calculation Report
6.3
Product Testing per EIA 541
EIA 541 does not require discharge pulse verification and does not specify a
specific bag size, but only requires that the capacitive sensor be placed 2" (5
cm) in from the open end of the bag and centered latterly. Therefore, the red
plastic adjustable stops should be inserted in the first set of holes for all testing.
As with the energy test, when the bag size increases, its capa