ETS 4431-EV Operating Manual

Text of ETS 4431-EV Operating Manual

SHIELDED BAG TEST SYSTEM

Model 4431-EV (Voltage & Energy) & Model 4431-V (Voltage Only) (These models replace Model 431) Operating Manual 10/10

(Equipment containing HV)

The equipment described in this Manual is designed and manufactured to operate within defined design limits. Any misuse may result in electric shock or fire. To prevent the equipment from being damaged, the following rules should be observed for installation, use and maintenance. Read the following safety instructions before operating the instrument. Retain these instructions in a safe place for fut

DO NOT OPERATE IN WET/DAMP CONDITIONS: If water or other liquid penetrates the

equipment, unplug the power cord and contact the factory for further instructions. Continuous use in this case may result in electrical shock, fire or permanent damage to the equipment. DO NOT OPERATE IN HIGH HUMIDITY: Operating the equipment in high humidity conditions will cause deteriation in performance, system failure, or present a shock or fire hazard. Contact the factor

The rapid advancement in the electronics industry during the past decade has placed

increasing importance on the understanding of electrostatics and its effect on electronic devices and systems. Electrostatic Discharge (ESD) is a common cause of microelectronic circuit failure. Many of these devices can be seriously damaged or destroyed by an electrostatic discharge below 30 Volts, or as a result of an electrostatic field of only a few h

2.0 TEST METHODS

2.1 ANSI/ESD STM11.31-2006 ESD Association standard for evaluating the performance of electrostatic discharge shielding materials - Bags This test method evaluates the performance of electrostatic discharge shielding bags. The purpose of this standard is to ensure that testing laboratories, using this test method to evaluate a given packaging material, will obtain similar results. This standard specifies the discharge waveform characteri

Figure 2.0-3: Typical Model 4431 shielded bag waveform

Six specimens of a given sample are required. Six measurements per specimen is specified. Testing is to be performed at both 12% and 50% ±3% RH at a temperature of 73 ±3°F after a conditioning period of 48 hours minimum. Bag size for this test should be 8x10 inches (20x25 cm). ANSI/ESD STM11.31 specifies that the test conditions, peak current, minimum, maximum and average energy levels of all 36 m

Figure 2.0-4: Voltage waveform for non-shielding and shielded bags

A good static shielding bag typically will have a voltage differential of 5 to 30 Volts. Bags with voltage differentials above 150 Volts usually have a surface resistance above the specified 1x103 Ohms upper limit. The voltage differential created by a static shielding bag is primarily caused by the time delay for the applied pulse to be induced on the lower electrode created by the resis

ETS 4431-EV Figure 3.0-1b: Model 4431-EV Shielded Bag Test System, laptop computer

This is a turn-key system that performs both the ANSI/ESD STM11.31 and the EIA 541-1998 shielding bag tests. Depending on the test method selected the System provides the correct discharge current pulse for both the ANSI/ESD STM11.31 and EIA 541 tests. The capacitive sensor detects the current pulse or the voltage differential and sends the signal(s) to the included oscilloscope for d

ETS 4431-EV 9. Adjustable bag insertion stops for configuring the System for different

bag sizes per specification. 10. Calibration jumpers and cables 1.1.2 Oscilloscope 1. Tektronix Model TDS2022B 3.1.2 Computer (Desktop or Laptop) 1. Running Windows� 2000/XP/VISTA/7 3.1.3 Software 1. ETS Test Suit Manager, Energy Calculation Program (Version 4.0.3.3) 3.2. Model 4431-V The Model 4431-V Discharge Unit is identical to the Model 4431-EV except it does not include th

a foot operated switch, which is convenient when testing in a humidity chamber. In the

AUTO mode, the discharge sequence specified in ANSI/ESD STM11.31 along with the scope settings are controlled automatically by the ETS Energy Calculation Program. The discharge waveform generated by the Model 4431 meets the requirements of ANSI/ESD STM5.1 Electrostatic Discharge Testing - Human Body Model. A typical discharge waveform measured with the discharge electr

5. USB to serial converter with 9-pin sub-D from computer to communications

cable (Model 4431-EV only) NOTE: Do not sustitute 6. 9-pin sub-D from Converter to 25-pin sub-D to Model 4431-EV 7. Foot switch 8. IEC Line Cord (North American plug) Figure 4.0-3b shows the jumpers and cables required to perform first time set up. Figure 4.0-3b: Calibration cables and jumpers for first time set up NOTE: If the Model 4431 is used to replace a Model 431 discharge

Figure 5.0-1: System set up for Energy test

1. Connect the BNC connector from the built-in CT-1 current transducer into CH-1 of the oscilloscope. The Terminator provided must be used in order to match the 50 Ohm output impedance of the CT-1 to the 1 MegOhm impedance of the scope. 2. Connect the 25 pin – 9-pin cable to the USB – 9-pin adapter and 25-pin sub- D connector on the rear of the Model 4431-EV 3. Connect the USB - USB cable to the 1st comput

Figure 5.0-2: Voltage differential test set up

If the Model 4431-EV is used to measure voltage differential, the CT-1 current sensor must be disconnected and the Voltage cables connected to perform this test. The Model 4431-EV is shipped from the factory with the capacitive sensor programmed for the Energy test. To convert the System from energy to voltage differential, the capacitive sensor must first be converted by plugging the red jumpers into the c

ETS 4431-EV 1 Megohm (10,000 and 10 MegOhms/sq). Current industry practice is to

use 1500 Ohms. 5.2.2 Oscilloscope The oscilloscope and Discharge Unit must first be calibrated as a set. If the oscilloscope is supplied as part of the system, the Discharge Unit and scope will have already been calibrated. If the scope is provided by the user, it is first necessary to compensate the built in 100:1 attenuators and cables to the scope input using the following procedure

Figure 5.0-4: Measured waveforms

Repeat the above for CH-2. The amplitude of each signal must be the same. If not, adjust the CH-1 and/or 2 “R” controls to obtain equal signals. After calibrating each channel individually, invert the CH-2 scope input and select ADD. The composite waveform should look like that shown in Figure 5.0-5c. Slight “R” and “C” adjustments may be required t

ETS 4431-EV especially if a digital storage oscilloscope is used, causing significant

measurement error. This condition normally occurs when buried metal layer bags are tested and 5, 10 or 20 Volt sensitivity levels are selected to obtain greater voltage differential resolution. To verify the signals, switch out of the ADD mode and into the CHOP mode and observe the individual CH-1 and CH-2 signals. d. Set the trigger to NORM and CH-1 trigger source. Adjust the lev

18 gage wire

Figure 6.0-1: Calibration set up to measure ESD pulse 6.1.2 EIA 541-1988 Voltage Differential EIA 541 does not specify the discharge waveform characteristics. However, it is implied that the waveform meets the requirements of ANSI/ESD STM 5.1. To verify the discharge waveform of the Model 4431- V, follow the procedure described in paragraph 4 in Section 5.1.1 above. 6.2 Product Testing per ANSI/ESD STM11.31 ANSI/ESD STM11.31

Set the oscilloscope vertical sensitivity to 50mV/div, time base to 50nsec/div and

trigger level to 50mV. Trigger mode is NORMAL. Turn on the HV. Initiate a discharge, using either the front panel TEST switch or the foot switch. The discharge pulse will activate for approximately 250 milliseconds. The current pulse detected by the capacitive sensor will be displayed on the oscilloscope. Typical waveforms for a static shielding bag are shown in Figure 6.

Figure: 6.0-4: Energy Calculation Report

6.3 Product Testing per EIA 541 EIA 541 does not require discharge pulse verification and does not specify a specific bag size, but only requires that the capacitive sensor be placed 2" (5 cm) in from the open end of the bag and centered latterly. Therefore, the red plastic adjustable stops should be inserted in the first set of holes for all testing. As with the energy test, when the bag size increases, its capa

ETS Test Equipment Owners Instructions: