ETS 220B Operating Manual

Text of ETS 220B Operating Manual

Electro-Tech Systems – 700 West Park Avenue, Perkasie, Pennsylvania 18944

D01654 Revision B – 2020-10-01 – Page 1 High Impedance Voltmeter ETS Model 220B Operating Manual D01654

Model 220B - High Impedance Voltmeter – Operating Manual

D01654 Revision - Page 2 of 15 Table of Contents I. Important Safety Information II. Description of contents III. Setup Guide IV. Quick Start Guide V. Functionality VI. Operation Details VII. Maintenance/Calibration VIII. Warranty IX. Appendix A - Computer Interface Revision History: Revision A: Legacy version 2018 Revision B: Released 2020-11-05. Reformat, add new address. Products des
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 3 of 15 I. IMPORTANT SAFETY INFORMATION SAFETY INSTRUCTIONS The equipment described in this Manual is designed and manufactured to operate within defined design limits. Any misuse may result in injury, electric shock, or discomfort to a person, incorrect readings, or damage to the equipment. The following rules should be observed for installation, use and maintenance. Read
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 4 of 15 II. Description of Contents Included: Item Qty. Description Model 220B 1 Circuitry in a metal enclosure with attached 42-inch probe cable with red clip probe. Ground Cable 1 1 A 42-inch black rubber cable featuring a black banana plug and an alligator clip Ground Cable 2 1 A 5’ black rubber cable featuring a green banana plug and a alligator clip Communication Cab
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 5 of 15 III. Set-Up Guide Step 1 – Unpack the unit Verify all parts are present and have survived shipping. Step 2 Initial Setup Connect the BLACK grounding cable to the banana jack (GND) on the front panel. If desired, the bail on the bottom of the Voltmeter case may be flipped down to tilt the Voltmeter upward toward the user, making it easier to read. Step 3 Power On
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 6 of 15 Step 4 - Initial Status Check A. If the BATT LO condition is on, replace the batteries (2x9V) before making any measurements. See the instructions in Section VI of this manual. B. Apply zero voltage input to the Voltmeter by clipping ground cable to HV probe as shown. Display should read 0 volts. C. With 0 volts applied, if the voltage meter does not read 0V, or i
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 7 of 15 IV. Quick Start Guide Step 1 – Turn on Power Turn on Power, verify display is active. Step 2 – Probe the Voltage First connect the ALLIGATOR CLIP on the grounding cable to a suitable ground (zero voltage) point. Then hold the RED HIGH VOLTAGE PROBE by its red plastic body and place the pointed metal tip in contact with the circuit to be measured. Step 3 – Disp
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 8 of 15 V. Functionality Introduction In many instances, the accurate measurement of D.C. voltages in excess of 1,000 Volts, which are produced by power supplies or voltage dividers with output (source) impedances in excess of 10 megohms, is required. In these situations, most readily available laboratory D.C. voltmeters will be useless since they will have insufficient ra
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 9 of 15 The digital meter displays the measure voltage directly in ±volts. The digital meter also provides a low battery indication to warn the user that the batteries should be changed. An over-scale indication warns the user that the input range of the unit has been exceeded. The circuits within the Model 220B are protected against over-scale inputs and external transie
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 10 of 15 VI. Operation of the Model 220B Display Details The meter contains a dot-matrix display readout that functions when power is on. When the high voltage input to the Voltmeter is within the nominal range of ±10,000VDC, the display will indicate the value of the voltage being measured directly in volts. The sign to the left of the number is the polarity of the input
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 11 of 15 Battery Replacement To replace the unit’s 9-Volt batteries, disconnect the high voltage probe from any voltage source being measured and turn off the power switch on the Meter’s front panel. On the rear panel is a 2-drawer battery compartment. Lift up each drawer and pull out. Remove the old batteries and insert the new ones. Be careful to observe the polarity.
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 12 of 15 Alternate Grounding Method If it is not possible (or not desirable) to use the ALLIGATOR CLIP on the grounding cable, the GND connection on the rear panel of the Voltmeter housing may be used to connect the Voltmeter to system ground (zero potential). Any suitable grounding wire may be used or the included GREEN CABLE. Over-Voltage Condition If the high voltage inp
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 13 of 15 VII. Maintenance and Calibration Calibration For consistency of performance Annual manufacturer conducted calibration is required. As with any test instrument, normal component aging and drift will require that the system be tested and calibrated by a qualified test and calibration facility on a periodic basis to insure that the unit is performing within its specif
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 14 of 15 VIII. Warranty Electro-Tech Systems, Inc. warrants its equipment, accessories and parts of its manufacture to be and remain free from defects in material and workmanship for a period of one (1) year from date of invoice and will, at the discretion of Seller, either replace or repair without charge, F.O.B. Perkasie, Pennsylvania, similar equipment or similar part t
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 15 of 15 Appendix A. Computer Interface The Model 220B has a 9-Pin sub min-D RS232A COMM PORT (Note 1) that provides real time data through serial communication. A PC will receive data by simply sending character R (for result) to the device. A simple program must be written by the user in order to collect data. The data can then be transferred to a spreadsheet or other pro