ETS 220B Operating Manual
Text of ETS 220B Operating Manual
Electro-Tech Systems – 700 West Park Avenue, Perkasie, Pennsylvania 18944
D01654 Revision B – 2020-10-01 – Page 1
High Impedance Voltmeter
ETS Model 220B Operating Manual
D01654
Model 220B - High Impedance Voltmeter – Operating Manual
D01654 Revision - Page 2 of 15
Table of Contents
I.
Important Safety Information
II. Description of contents
III. Setup Guide
IV. Quick Start Guide
V. Functionality
VI. Operation Details
VII. Maintenance/Calibration
VIII. Warranty
IX. Appendix A - Computer Interface
Revision History:
Revision A: Legacy version 2018
Revision B: Released 2020-11-05. Reformat, add new address.
Products des
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 3 of 15
I.
IMPORTANT SAFETY INFORMATION
SAFETY INSTRUCTIONS
The equipment described in this Manual is designed and manufactured to operate within defined
design limits. Any misuse may result in injury, electric shock, or discomfort to a person, incorrect
readings, or damage to the equipment. The following rules should be observed for installation, use
and maintenance. Read
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 4 of 15
II. Description of Contents
Included:
Item
Qty.
Description
Model 220B
1
Circuitry in a metal enclosure with attached 42-inch probe cable
with red clip probe.
Ground Cable 1
1
A 42-inch black rubber cable featuring a black banana plug and an
alligator clip
Ground Cable 2
1
A 5’ black rubber cable featuring a green banana plug and a
alligator clip
Communication Cab
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 5 of 15
III. Set-Up Guide
Step 1 – Unpack the unit
Verify all parts are present and have survived
shipping.
Step 2 Initial Setup
Connect the BLACK grounding cable to the
banana jack (GND) on the front panel.
If desired, the bail on the bottom of the
Voltmeter case may be flipped down to tilt the
Voltmeter upward toward the user, making it
easier to read.
Step 3 Power On
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 6 of 15
Step 4 - Initial Status Check
A. If the BATT LO condition is on, replace
the batteries (2x9V) before making any
measurements. See the instructions in
Section VI of this manual.
B. Apply zero voltage input to the Voltmeter
by clipping ground cable to HV probe as
shown. Display should read 0 volts.
C. With 0 volts applied, if the voltage meter
does not read 0V, or i
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 7 of 15
IV. Quick Start Guide
Step 1 – Turn on Power
Turn on Power, verify display is active.
Step 2 – Probe the Voltage
First connect the ALLIGATOR CLIP on the
grounding cable to a suitable ground (zero
voltage) point. Then hold the RED HIGH
VOLTAGE PROBE by its red plastic body
and place the pointed metal tip in contact
with the circuit to be measured.
Step 3 – Disp
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 8 of 15
V. Functionality
Introduction
In many instances, the accurate measurement of D.C. voltages in excess of 1,000 Volts, which
are produced by power supplies or voltage dividers with output (source) impedances in excess of
10 megohms, is required. In these situations, most readily available laboratory D.C. voltmeters
will be useless since they will have insufficient ra
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 9 of 15
The digital meter displays the measure voltage directly in ±volts. The digital meter also provides a
low battery indication to warn the user that the batteries should be changed. An over-scale
indication warns the user that the input range of the unit has been exceeded. The circuits within
the Model 220B are protected against over-scale inputs and external transie
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 10 of 15
VI. Operation of the Model 220B
Display Details
The meter contains a dot-matrix display
readout that functions when power is on.
When the high voltage input to the Voltmeter is
within the nominal range of ±10,000VDC, the
display will indicate the value of the voltage
being measured directly in volts. The sign to
the left of the number is the polarity of the input
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 11 of 15
Battery Replacement
To replace the unit’s 9-Volt batteries,
disconnect the high voltage probe from any
voltage source being measured and turn off the
power switch on the Meter’s front panel. On the
rear panel is a 2-drawer battery compartment.
Lift up each drawer and pull out. Remove the
old batteries and insert the new ones. Be
careful to observe the polarity.
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 12 of 15
Alternate Grounding Method
If it is not possible (or not desirable) to use
the ALLIGATOR CLIP on the grounding
cable, the GND connection on the rear
panel of the Voltmeter housing may be
used to connect the Voltmeter to system
ground (zero potential). Any suitable
grounding wire may be used or the
included GREEN CABLE.
Over-Voltage Condition
If the high voltage inp
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 13 of 15
VII. Maintenance and Calibration
Calibration
For consistency of performance Annual manufacturer conducted calibration is required.
As with any test instrument, normal component aging and drift will require that the system be
tested and calibrated by a qualified test and calibration facility on a periodic basis to insure
that the unit is performing within its specif
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 14 of 15
VIII. Warranty
Electro-Tech Systems, Inc. warrants its equipment, accessories and parts of its manufacture to be
and remain free from defects in material and workmanship for a period of one (1) year from date of
invoice and will, at the discretion of Seller, either replace or repair without charge, F.O.B. Perkasie,
Pennsylvania, similar equipment or similar part t
Model 220B - High Impedance Voltmeter – Operating Manual D01654 Revision - Page 15 of 15
Appendix A. Computer Interface
The Model 220B has a 9-Pin sub min-D RS232A COMM PORT (Note 1) that provides real time data through
serial communication. A PC will receive data by simply sending character R (for result) to the device. A simple
program must be written by the user in order to collect data. The data can then be transferred to a spreadsheet
or other pro