Bante Instruments TB100 Operation & User’s Manual
Text of the manual
1 I Overview
Introduction
Thank you for selecting the TB100 portable turbidity meter. This user
manual provides a step-by-step guide to help you operate the meter,
please carefully read the following instructions before use. Any use
outside of these instructions may invalidate your warranty and cause
permanent damage to the meter.
Environmental Conditions
Overview I 2
Display
Icon
Description
Indicates that the meter is in the measurement mode.
Indicates that the meter is in the calibration mode.
Indicates that the meter is in the setup mode.
Indicates that you are viewing the stored readings or a
reading is stored into the memory.
Indicates that the meter is calibrating.
Shows the battery status. Replace t
3 I Setup
Meter Setup
The TB100 turbidity meter contains 8 menu items in the setup menu,
the following table describes the functions of each option.
Menu Item
Option and Description
Calibration Points
Set the number of calibration points.
2 to 5 points (default 2 points)
Date and Time
Set the date and time for data logs.
Measurement Unit
Set the default tu
Calibration I 4
Measurement and Calibration Hints
• Do not hold the meter in hand during calibration and measurement.
• Never pour liquid directly into the measurement chamber.
• Keep the outside of vial clean and dry. If necessary, wipe the vial
with a lint-free cloth.
• If the vial has scratches or scuffs, replace a new one.
• Ensur
5 I Measurement
Measurement
Selecting the Measurement Mode
The TB100 turbidity meter contains two measurement modes. If the
single measurement mode is enabled, the meter will automatically
sense a stable reading and lock the measurement. If the continuous
measurement mode is enabled, the meter will continuously measure
and update readings.
1.
In the measur
Data Management I 6
Continuous Measurement
3.1 Ensure that the meter is in the continuous measurement mode.
3.2 Repeat steps 1.1 through 1.3 above.
3.3 Press the Meas key, the Measure icon begins flashing, the meter
continuously measures the sample.
To stop the measurement, press the Meas key again.
Data Management
Storing a Measurement Result
The TB100 tu
7 I Appendix
Appendix
Indexing a Sample Vial
The United States Environmental Protection Agency recommends that
the vial used for turbidity calibration or sample measurement be
indexed. Its purpose is to obtain a position which provides the lowest
turbidity reading. The indexing methods are as follows.
1.1 Fill the vial with the distilled water (< 0.5 NTU).
The information in this document is subject to change without notice.
Copyright © Bante Instruments Inc, 2022. All rights reserved.
Disposal