Bante Instruments TB100 Operation & User’s Manual

Text of the manual

1 I Overview

Introduction Thank you for selecting the TB100 portable turbidity meter. This user manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Any use outside of these instructions may invalidate your warranty and cause permanent damage to the meter. Environmental Conditions

Overview I 2

Display Icon Description Indicates that the meter is in the measurement mode. Indicates that the meter is in the calibration mode. Indicates that the meter is in the setup mode. Indicates that you are viewing the stored readings or a reading is stored into the memory. Indicates that the meter is calibrating. Shows the battery status. Replace t

3 I Setup

Meter Setup The TB100 turbidity meter contains 8 menu items in the setup menu, the following table describes the functions of each option. Menu Item Option and Description Calibration Points Set the number of calibration points. 2 to 5 points (default 2 points) Date and Time Set the date and time for data logs. Measurement Unit Set the default tu

Calibration I 4

Measurement and Calibration Hints • Do not hold the meter in hand during calibration and measurement. • Never pour liquid directly into the measurement chamber. • Keep the outside of vial clean and dry. If necessary, wipe the vial with a lint-free cloth. • If the vial has scratches or scuffs, replace a new one. • Ensur

5 I Measurement

Measurement Selecting the Measurement Mode The TB100 turbidity meter contains two measurement modes. If the single measurement mode is enabled, the meter will automatically sense a stable reading and lock the measurement. If the continuous measurement mode is enabled, the meter will continuously measure and update readings. 1. In the measur

Data Management I 6

Continuous Measurement 3.1 Ensure that the meter is in the continuous measurement mode. 3.2 Repeat steps 1.1 through 1.3 above. 3.3 Press the Meas key, the Measure icon begins flashing, the meter continuously measures the sample. To stop the measurement, press the Meas key again. Data Management Storing a Measurement Result The TB100 tu

7 I Appendix

Appendix Indexing a Sample Vial The United States Environmental Protection Agency recommends that the vial used for turbidity calibration or sample measurement be indexed. Its purpose is to obtain a position which provides the lowest turbidity reading. The indexing methods are as follows. 1.1 Fill the vial with the distilled water (< 0.5 NTU).

The information in this document is subject to change without notice.

Copyright © Bante Instruments Inc, 2022. All rights reserved. Disposal This product is required to comply with the European Union’s Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC and may not be disposed of in domestic waste. Please dispose of product in accordance

More Bante Instruments Measuring Instruments Owners Instructions: