Bante Instruments PHscan20 Instruction Manual

Text of the manual

Bante is committed to the reduction and eventual elimination of all hazardous

substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like. New technologies and design parameters are a

The tester includes 3 multifunction keys, names and symbols describe function

key controls. The tester is equipped with a clear LCD display used to show measured values, mode indicators and unit indicators. Thank you for selecting the PHscan20 pocket pH tester. This manual provides a step-by-step guide to help you operate t

— Press MEAS key to turn on the tester, the display shows measured values.

— — Preparation of pH Standard Buffer Solutions The tester is packaged with three pH buffer packets required for calibration. Open the pH7.00 buffer packet, place the powder into a 250ml volumetric

Press ENTER key to confirm, the tester goes to calibration point selection

mode, the display shows “2/CAL” (2 4. Press CAL key to select the number of calibration points. Press ENTER key to confirm, the 3. 5. tester shows “YES/HOLD” the auto-hold function is enabled. — If the auto-hold function is

POCKET PH TESTER

allows up to 3 points calibration. tester can — — — Press CAL key the tester shows “pH7.00/CAL1”(or “pH6.86/CAL1”). 4. Dip the pH electrode into the pH7.00 (or pH6.86) standard buffer solution, the end of the electrode must be completely immersed into the calibration solution. Stir the tes

Rinse the electrode with distilled water,

ait for the measured value to stabilize, the display shows electrode slope and “END”. The tester returns to the pH measurement mode. Calibration is completed. EXIT THE CALIBRATION: During the calibration, if you want to exit calibration mode, press MEAS key. The tester will i

1 to 3 points

Calibration Solutions USA (pH4.01, 7.00, 10.01) NIST (pH4.01, 6.86, 9.18) Temperature Compensation o o 0~60 C, 32~140 F, Automatic Auto-Off After 8 minutes from last key press Operating Temperature o o 0~60 C, 32~140 F LCD Display Custom LCD Power Requirement 2 1.5V “ × AAA” Batteries Weight 100 grams Dimensi