Q U I C K R E F E R E N C E G U I D E
Oxygen Assist Module
for Vapotherm Precision Flow® Systems
Alarm Silence
LED
Alarm Silence
Button
Setting Control
Knob
Touchscreen
Display
On/Off LED
On/Off Button
This guide provides you with basic instructions on how to set up and
operate the Vapotherm Oxygen Assist Module for Precision Flow®
Systems. Before operat
%O2 Above Limit*
Increase %O2 Alarm Limit, or enter MANUAL mode
Battery Backup In Use
Oxygen Assist Module is in Battery Mode. When external power is lost,
the module will automatically switch to Battery Mode
Battery Low
Plug the Oxygen Assist Module into external power
Check Precision Flow
Clear Precision Flow alarm and ensure it
INDICATIONS, WARNINGS AND CAUTIONS
Primary Indications:
The Oxygen Assist Module is an optional module used only with the Vapotherm Precision Flow and
is indicated for on-demand titration of oxygen into warm humidified breathing gases delivered to
spontaneously breathing patients based on continuous non-invasive monitoring of blood oxygen
saturation.
The Oxy
100 Domain Drive
Exeter, NH 03833
T: 603-658-0011
USA
3101889 Rev B
May be patented. www.vapotherm.com/patents
Technical Support
T: 855 557 8276 USA
+1 (603) 658-5121 International
[email protected]
AJW Technology Consulting GmbH
Königsallee 106
40215 Düsseldorf
Germany
Phone: +49 (0) 211 3013 2232
0297
SHUT DOWN
GENERAL WARNINGS
•
For use on