SCI 6000 Series Quick Start Manual
Text of SCI 6000 Series Quick Start Manual
6-in-1 MULTI-FUNCTION TESTER
QUICK START GUIDE
Models: 6330
SAFETY CHECKLIST
Survey the test station. Make sure it is safe & orderly.
Always keep unqualified/unauthorized personnel away from the test area.
Familiarize yourself with safety protocols in the event of a problem.
Exercise caution and never touch products or connections during a test.
Train operators. Connect the return lead first and never touch clips directly.
You should
SAFETY MADE SIMPLE®
2
Front Panel Controls
Reset Button: Red momentary contact switch used to reset the instrument in case of a
failure. Also serves as an abort signal to stop any test in progress.
Test Button: Starts a test.
LCD Display: 20 x 2 character display.
Lock Indicator: This LED indicates that the lock feature has been enabled.
Bus Interface Indicator: This LED indicates that the bus interface has been enabled.
Fault Condition Indicator: These L
Back Panel Controls
External Measuring Device: Provides an access port for a custom measuring device. The port
may be activated during Line Leakage testing from the instrument’s menu system. The operator
must supply the appropriate measuring device as none is provided.
Power Control
Fuse Receptacle: To change the fuse, unplug the power (mains) cord and turn the fuse
receptacle counterclockwise. The fuse compartment will be exposed. Please rep
SAFETY MADE SIMPLE® 4
1. Adjust the line voltage select switch on the rear panel of the instrument to the appropriate
input voltage rating, either 115 VAC or 230 VAC.
Tester Setup
WARNING: LOCATE A SUITABLE TESTING AREA WITH A THREE-PRONG, GROUNDED OUTLET. BE SURE THAT
YOUR THREE-PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ THE SAFETY CHECKLIST OF
THIS GUIDE BEFORE STARTING TO TEST.
2. Connect the female end of the standard NEMA style line
Configure the instrument system parameters to your preferences. The instrument system
parameters are global and will affect all tests that are performed regardless of memory
location and memory step.
Selecting a Memory Location
SELECTING A MEMORY LOCATION
SET
EXIT
SET
EXIT
SET
EXIT
Edit System Parameters
1. Press the SET key from the
Perform Test screen to select
a memory location. Use the
up or down arrow keys to
se
SAFETY MADE SIMPLE® 6
Edit Test Parameters
EDITING TEST PARAMETERS
SET
EXIT
SET
EXIT
SET
EXIT
SET
EXIT
SET
EXIT
1. Press the SET key from the
Perform Test screen to
confirm the memory
location.
EDITING TEST PARAMETERS
SET
EXIT
SET
EXIT
SET
EXIT
SET
EXIT
SET
EXIT
2. Press the SET key again to
confirm the memory step
number. You should start by
modifying step 1.
3. Press the SET key a third
time u
Default test test parameters for each test type are as follows:
Line
1
2
3
4
5
6
7
8
Neutral
OPEN
OPEN
OPEN
OPEN
CLOSE
CLOSE
CLOSE
CLOSE
Reverse
OFF
ON
OFF
ON
OFF
ON
OFF
ON
Ground
OPEN
OPEN
CLOSE
CLOSE
OPEN
OPEN
CLOSE
CLOSE
AC Hipot Mode
• Voltage: 1.24 kV
• HI-Limit: 10.00 mA
• Lo-Limit: 0.00 mA
• Ramp: 0.1 s
• Dwell: 1.0 s
• Frequency: 60 Hz
• DUT Output: On
• Connect: Off
DC Hipot Mode
• Voltage: 1.50 kV
• HI-Limit: 5.00