SCI 6000 Series Quick Start Manual

Text of SCI 6000 Series Quick Start Manual

6-in-1 MULTI-FUNCTION TESTER

QUICK START GUIDE Models: 6330 SAFETY CHECKLIST Survey the test station. Make sure it is safe & orderly. Always keep unqualified/unauthorized personnel away from the test area. Familiarize yourself with safety protocols in the event of a problem. Exercise caution and never touch products or connections during a test. Train operators. Connect the return lead first and never touch clips directly. You should

SAFETY MADE SIMPLE®

2 Front Panel Controls Reset Button: Red momentary contact switch used to reset the instrument in case of a failure. Also serves as an abort signal to stop any test in progress. Test Button: Starts a test. LCD Display: 20 x 2 character display. Lock Indicator: This LED indicates that the lock feature has been enabled. Bus Interface Indicator: This LED indicates that the bus interface has been enabled. Fault Condition Indicator: These L

Back Panel Controls

External Measuring Device: Provides an access port for a custom measuring device. The port may be activated during Line Leakage testing from the instrument’s menu system. The operator must supply the appropriate measuring device as none is provided. Power Control Fuse Receptacle: To change the fuse, unplug the power (mains) cord and turn the fuse receptacle counterclockwise. The fuse compartment will be exposed. Please rep
SAFETY MADE SIMPLE® 4 1. Adjust the line voltage select switch on the rear panel of the instrument to the appropriate input voltage rating, either 115 VAC or 230 VAC. Tester Setup WARNING: LOCATE A SUITABLE TESTING AREA WITH A THREE-PRONG, GROUNDED OUTLET. BE SURE THAT YOUR THREE-PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST. 2. Connect the female end of the standard NEMA style line

Configure the instrument system parameters to your preferences. The instrument system

parameters are global and will affect all tests that are performed regardless of memory location and memory step. Selecting a Memory Location SELECTING A MEMORY LOCATION SET EXIT SET EXIT SET EXIT Edit System Parameters 1. Press the SET key from the Perform Test screen to select a memory location. Use the up or down arrow keys to se
SAFETY MADE SIMPLE® 6 Edit Test Parameters EDITING TEST PARAMETERS SET EXIT SET EXIT SET EXIT SET EXIT SET EXIT 1. Press the SET key from the Perform Test screen to confirm the memory location. EDITING TEST PARAMETERS SET EXIT SET EXIT SET EXIT SET EXIT SET EXIT 2. Press the SET key again to confirm the memory step number. You should start by modifying step 1. 3. Press the SET key a third time u

Default test test parameters for each test type are as follows:

Line 1 2 3 4 5 6 7 8 Neutral OPEN OPEN OPEN OPEN CLOSE CLOSE CLOSE CLOSE Reverse OFF ON OFF ON OFF ON OFF ON Ground OPEN OPEN CLOSE CLOSE OPEN OPEN CLOSE CLOSE AC Hipot Mode • Voltage: 1.24 kV • HI-Limit: 10.00 mA • Lo-Limit: 0.00 mA • Ramp: 0.1 s • Dwell: 1.0 s • Frequency: 60 Hz • DUT Output: On • Connect: Off DC Hipot Mode • Voltage: 1.50 kV • HI-Limit: 5.00