Extech Instruments MA400 Operation & User’s Manual
Text of Extech Instruments MA400 Operation & User’s Manual
MA Series V2.0 7/02
2
WARRANTY
EXTECH
INSTRUMENTS
CORPORATION
warrants
this
instrument to be free of defects in parts and workmanship for one
year from date of shipment (a six month limited warranty applies
on sensors and cables). If it should become necessary to return
the instrument for service during or beyond the warranty period,
contact the Customer Service Department at (781) 890-7440 ext.
210 for authorization. A Return Authorization (RA) number m
MA Series V2.0 7/02 3
Introduction
Congratulations on your purchase of Extech model MA400,
MA420, MA440 or MA460+ clamp-on multimeter. Properly used,
this meter will provide many years of reliable service.
Clamp meter functions common to all meters:
Autoranging
AC Current Measurements
Data Hold
Auto Power Off
Maximum Hold captures 30ms peaks
1.1" (28mm) Jaw size
MultiMeter functions common to all models:
AC/DC Voltage to 600V
Resistance to 40MΩ
Capacit
MA Series V2.0 7/02 4
Safety Precautions
1.
Improper use of this meter can cause damage, shock,
injury or death. Read and understand this users manual
before operating the meter.
2.
Make sure any covers or battery doors are properly closed
and secured.
3.
Always remove the test leads before replacing the battery
or fuses.
4.
Inspect the condition of the test leads and the meter itself
for any damage before operating the meter. Repair or
replace any damag
MA Series V2.0 7/02 5
PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY
OVERVOLTAGE CATEGORY I
Equipment of OVERVOLTAGE CATEGORY I is equipment for
connection to circuits in which measures are taken to limit the
transient overvoltages to an appropriate low level. Note –
Examples include protected electronic circuits.
OVERVOLTAGE CATEGORY II
Equipment of OVERVOLTAGE CATEGORY II is energy-
consuming equipment to be supplied from the fixed installation.
Note
MA Series V2.0 7/02 6
Specifications
DC Voltage
RANGE
Accuracy
400.0mV
0.3% + 4d
4.000V, 40.00V, 400.0V
0.5% + 3d
600V
1.0% + 4d
NMRR: >50dB @ 50/60Hz, CMRR: >120dB @ DC, 50/60Hz,
Rs=1kΩ
Input impedance: 10MΩ, 30pF nominal, 1000MΩ on the 400.0mV
range.
AC Voltage
RANGE (50Hz to 500Hz)
Accuracy
4.000V, 40.00V, 400.0V
1.5% + 5d
600V
2.0% + 5d
CMRR: >60dB @ DC to 60Hz, Rs=1kΩ, Input Impedance:
10MΩ, 30pF nominal, True RMS models specified from 5%
MA Series V2.0 7/02 7
Capacitance
RANGE
Accuracy*
500.0nF, 5.000µF, 50.00µF, 500.0µF,
3000µF
3.5%* + 6d
50.00nF
Not specified
*Accuracies with a film capacitor or better
*Specified with battery voltage above 2.8V (approximately half full
battery). Accuracy decreases gradually to 12% at low battery
warning voltage of approximately 2.4V.
Frequency
Function
Sensitivity
(ACrms)
Range
400.0mV
150mV
10Hz to 2kHz
4.000V
3.2V
5Hz to 40kHz
40.00V
25V
5Hz to 1
MA Series V2.0 7/02 8
Accuracy Notes: Accuracy is ± (% reading digits + number of
digits), or as otherwise specified, at 23°C ±5°C < 75% R.H.
Model MA420 True RMS accuracies are specified from 10%
to 100% of range or as otherwise specified. Maximum Crest
Factor <1.75:1 at full scale & <3.5:1 at half scale (with
frequency component within the specified frequency
bandwidth for non-sinusoidal waveforms).
Audible Continuity:
Threshold between 10Ω and 1
MA Series V2.0 7/02 9
Safety:
The MiniAmp™ Series meters meet
IEC61010-032(1994), EN61010-2-
032(1995), UL3111-2-032(1999).
MA440 and MA460+: Category III
600V AC & DC
MA400 and MA420: Category II 600V
AC & DC and Category III 300V AC &
DC
E.M.C.:
Meets EN61326(1997, 1998/A1),
EN61000-4-2(1995), and EN61000-4-
3(1996)
In an RF field of 1V/m:
Capacitance function is not specified
µA function: Total accuracy = Acc.+50
digits
Other functions: Total acc.
MA Series V2.0 7/02 10
Meter Description
MA440 and MA460+ (double display models)
1.
Clamp jaw
2.
Jaw trigger
3.
ACA LCD display
4.
Multimeter LCD display
5.
Select button
6.
Multimeter function switch
7.
Multimeter input jacks
8.
Hz button
9.
Clamp power button
10. Data Hold/Max Hold button
MA400 and MA420 (single display models)
1.
Clamp jaw
2.
Jaw trigger
3.
LCD display
4.
Select button
5.
Multimeter function switch
6.
Multimeter input jacks
7.
Hz but
MA Series V2.0 7/02 11
Operation
NOTICE: Read and understand all warning and caution
statements listed in the safety section of this operation manual
prior to using this meter.
AC/DC Voltage Measurements
1.
Connections are made through the test lead terminals.
2.
Insert the black test lead into the negative COM jack and
the red test lead into the positive jack.
3.
Slide the function switch to the "V" position. Press the
SELECT key momentarily to togg
MA Series V2.0 7/02 12
Resistance and Continuity Measurements
1.
Insert the black test lead banana plug into the negative
COM jack and the red test lead banana plug into the
jack.
2.
Slide the function switch to the "Ω •)))" position.
3.
Press the SELECT key momentarily to select Continuity.
4.
Touch the test probe tips across the circuit or part under
test. It is best to disconnect one side of the part under test
so the rest of the circuit will not
MA Series V2.0 7/02 13
electrical noise that may produce an unstable reading. However,
if the HZ reading displays zero due to insufficient sensitivity,
activate the HZ function before measuring the signal to select
the highest Hz sensitivity.
Diode Test
1.
Insert the black lead into the negative COM jack and the
red test lead into the positive jack
2.
Slide the function switch to " " position.
3.
Press the SELECT key twice until “ “ and “
MA Series V2.0 7/02 14
Max Hold
1.
Press and hold for 1 second the Max Hold key to capture
and hold the maximum measured value (up to 30ms). The
annunciators “H MAX” will appear in the display
2.
Press and hold for 1 second the Max Hold key to exit the
mode.
Note: The Max Hold feature applies only to the upper current
display (ACA) on the MA440 and MA460+ double display
models. It applies to the ACA, DCV and ACV functions on
the single display MA400
MA Series V2.0 7/02 15
Tech Support Hotlines
781-890-7440 ext. 200
[email protected]
www.extech.com
Maintenance
WARNING
To avoid electrical shock, disconnect the meter from any circuit,
remove the test leads from the input jacks and turn OFF the meter
before opening the case. Do not operate with open case.
Cleaning and Storage
Periodically wipe the case with a damp cloth and mild detergent;
do not use abrasives or solvents. If the meter is not to be used