Extech Instruments MN16A Operation & User’s Manual

Text of Extech Instruments MN16A Operation & User’s Manual

MN16A V1.4 03/12

2 Introduction Congratulations on your purchase of the Extech MN16A Autoranging Multimeter. This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency, Diode Test, and Continuity plus Thermocouple Temperature. It features a rugged design for heavy duty use This meter is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Safety This symbol adjacent to another symbol
MN16A V1.4 03/12 3 CAUTIONS • Improper use of this meter can cause damage, shock, injury or death. Read and understand this user manual before operating the meter. • Always remove the test leads before replacing the battery or fuses. • Inspect the condition of the test leads and the meter itself for any damage before operating the meter. Repair or replace any damage before use. • Use great care when making measurements if the voltages are greate
MN16A V1.4 03/12 4 SAFETY INSTRUCTIONS This meter has been designed for safe use, but must be operated with caution. The rules listed below must be carefully followed for safe operation. 1. NEVER apply voltage or current to the meter that exceeds the specified maximum: Input Protection Limits Function Maximum Input V DC or V AC 600VDC/AC mA AC/DC 400mA AC/DC A AC/DC 10A AC/DC (20A for 30 sec max every 15 min) Frequency, Resistance, Capacitance, Diode Te
MN16A V1.4 03/12 5 Controls and Jacks 1. 4,000 count LCD display 2. MODE button 3. HOLD button 4. Function switch 5. COM input jack 6. 10A input jack 7. V, Ω, CAP, Temp, Hz, μA and mA input jack 8. REL button 9. RANGE button 10. Battery compartment (rear) Symbols and Annunciators •))) Continuity Diode test Low Battery n nano (10 -9) (capacitance) µ micro (10 -6) (amps) m milli (10 -3) (volts, amps) A Amps k kilo (10 3) (ohms) F Farads (capacitance
MN16A V1.4 03/12 6 Operating Instructions WARNING: Risk of electrocution. High-voltage circuits, both AC and DC, are very dangerous and should be measured with great care. 1. ALWAYS turn the function switch to the OFF position when the meter is not in use. 2. If “OL” appears in the display during a measurement, the value exceeds the range you have selected. Change to a higher range. DC VOLTAGE MEASUREMENTS CAUTION: Do not measure DC voltages if a m
MN16A V1.4 03/12 7 DC CURRENT MEASUREMENTS CAUTION: Do not make 20A current measurements for longer than 30 seconds. Exceeding 30 seconds may cause damage to the meter and/or the test leads. 1. Insert the black test lead banana plug into the negative COM jack. 2. For current measurements up to 4000µA DC; Set the function switch to the µA position and insert the red test lead banana plug into the µA jack. 3. For current measurements up to 400mA DC; S
MN16A V1.4 03/12 8 RESISTANCE MEASUREMENTS WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any resistance measurements. Remove the batteries and unplug the line cords. 1. Set the function switch to the Ω position. 2. Insert the black test lead banana plug into the negative COM jack. Insert the red test lead banana plug into the positive Ω jack. 3. Press the MODE button to indicate �
MN16A V1.4 03/12 9 TEMPERATURE MEASUREMENTS 1. Set the function switch to the ºF or ºC position. 2. Insert the Temperature Probe into the input jacks, making sure to observe the correct polarity. 3. Touch the Temperature Probe head to the part whose temperature you wish to measure. Keep the probe touching the part under test until the reading stabilizes (about 30 seconds). 4. Read the temperature in the display. Note: The temperature probe is fitted w
MN16A V1.4 03/12 10 AUTORANGING/MANUAL RANGE SELECTION When the meter is first turned on, it automatically goes into Autoranging. This automatically selects the best range for the measurements being made and is generally the best mode for most measurements. For measurement situations requiring that a range be manually selected, perform the following: 1. Press the RANGE button. The “Auto” display indicator will turn off. 2. Press the RANGE key to ste
MN16A V1.4 03/12 11 Maintenance WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the back cover or the battery or fuse covers. WARNING: To avoid electric shock, do not operate your meter until the battery and fuse covers are in place and fastened securely. This MultiMeter is designed to provide years of dependable service, if the following care instructions are performed: 1. KEEP THE METER DRY. If i
MN16A V1.4 03/12 12 Specifications Function Range Resolution Accuracy 4V 0.001V 40V 0.01V 400V 0.1V ±(1.2% reading + 2 digits) DC Voltage (V DC) 600V 1V ±(1.5% reading + 2 digits) 400mV* 0.1mV ±(1.5% reading + 15 digits) 4V 0.001V ±(1.2% reading + 3 digits) 40V 0.01V 400V 0.1V ±(1.5% reading + 3 digits) 600V 1V ±(2.0% reading + 4 digits AC Voltage (V AC) (50 / 60Hz) *400mV range is not autoranging 400μA 0.1μA ±(1.0% reading + 3 digits) 4000μA
MN16A V1.4 03/12 13 Function Range Resolution Accuracy 5.000Hz 0.001Hz 50.00Hz 0.01Hz ±(1.5% reading + 5 digits) 500.0Hz 0.1Hz 5.000kHz 0.001kHz 50.00kHz 0.01kHz 500.0kHz 0.1kHz ±(1.2% reading + 3 digits) 5.000MHz 0.001MHz ±(1.5% reading + 4 digits) Frequency Sensitivity: >0.5V rms min. [ 1MHz; >3V rms > 1MHz NOTE: Accuracy is stated at 65 oF to 83 oF (18 oC to 28 oC) and less than 75% RH. Diode Test Test current of 0.3mA typical, open circuit voltag
MN16A V1.4 03/12 14 Support lines: U.S. 877-439-8324, Intl. 603-324-7800 Technical support: Option 3; E-mail: [email protected] Repair & Returns: Option 4; E-mail: [email protected] Product specifications subject to change without notice Visit our website: www.extech.com Extech Instruments Corporation, 9 Townsend West, Nashua, NH 03063 ISO 9001 Certified since 1995 Warranty EXTECH INSTRUMENTS CORPORATION (a FLIR company) warrants this instrument to be

Extech Instruments Multimeter Owners Instructions: