Extech Instruments 407227 Operation & User’s Manual

Text of Extech Instruments 407227 Operation & User’s Manual

Heavy Duty pH, mV, & Temperature Meter

Model 407227 • Measures pH, mV and Temperature • RS-232 PC Interface • oC/ oF switchable • Front panel Slope and Calibration adjustments Introduction Congratulations on your purchase of Extech’s pH, mV, and Temperature Meter. This professional meter, with proper care, will provide years of safe reliable service. Shop for Extech products at: 1.877.766.5412 ShopExtech.com

Model 407227 Ver. 2.1 Oct. 2002

Page 2 Specifications General Specifications Display Dual function LCD with contrast adjust Input impedance 1012 ohms Temp. compensation for pH Automatic and manual 32 to 149°F (0 to 65°C) pH electrode compatibility Any combination pH electrode with BNC connector Data hold Freezes displayed reading Memory recall Records / Recalls MAX (maximum), MIN (minimum), and AVG (average) readings Auto Power OFF Meter powers off a
Model 407227 Ver. 2.1 Oct. 2002 Page 3 Temperature Compensation For accurate pH measurements and calibration, the temperature of the solution under test must be known. This can be accomplished manually or automatically. Read through the steps for each (below) and determine that best method for the application at hand: Automatic Temperature Compensation NOTE: For Automatic Compensation, the supplied thermometer must be used. Perform these steps before e
Model 407227 Ver. 2.1 Oct. 2002 Page 4 pH Operation IMPORTANT NOTE: Manual or Automatic Temperature Compensation (covered earlier in this manual) must be performed before proceeding with pH calibration and pH measurements. pH CALIBRATION Note that the meter must be calibrated before each use 1. Press the POWER key to turn the meter on (if it isn’t already on). 2. Connect the Combination pH electrode to the meter’s BNC input socket and place the ele
Model 407227 Ver. 2.1 Oct. 2002 Page 5 Advanced Features MIN MAX AVG Record / Recall Mode 1. Press the RECORD key (the REC indicator will appear on the LCD). 2. Press the CALL key to view the highest (MAX) reading recorded since the RECORD key was first pressed. The MAX indicator will appear on the LCD. 3. Press the CALL key to view the lowest (MIN) reading recorded since the RECORD key was first pressed. The MIN indicator will appear on the LCD. 4. Pr
Model 407227 Ver. 2.1 Oct. 2002 Page 6 Support Hotline (781) 890-7440 Tech support: Ext. 200; Email: [email protected] Repair/Returns: Ext. 210; Email: [email protected] Website: www.extech.com Battery Replacement The low battery indicator (LBT) appears on the LCD when the 9V battery needs replacing. 1. Remove the meter's rubber protective holster. 2. Open the battery compartment using a small coin or screwdriver. 3. Replace the 9V battery and re-asse

Extech Instruments Multimeter Owners Instructions: