All Rights Reserved, Copyright  1995, RKC INSTRUMENT INC. Resin Pressure Measuring System Instruction Manual CZ-100P (Resin Pressure Sensor) PCT-300 (Output Converter) IM100CZ04-E2 Thank you for purchasing this RKC product. In order to achieve maximum performance and ensure proper operation of your new instrument, carefully read all ...

1.1 Principle of Operation

When pressure is applied to a diaphragm in the resin pressure sensor (CZ-100P), force acts upon a pressure sensing element through a metal rod located near the diaphragm. The pressure sensing element used in CZ-100P has 4 gauges adhered to a metallic elastic body and connected to Wheatstone bridge. Each st ...

• Do not use the pressure sensor under any of the following

environmental conditions: - Where the sensor is exposed directly to cold air, warm air or hot air. - Where temperature variations are large. - Where the sensor is exposed to direct sunlight. - Where the sensor is directly splashed with water or rain, or the humidity is high. • Do ...

• A lead pipe cover (option of HB and HC type) is mainly for protecting

the exposed section below the sensor outer cylinder from being exposed to cold wind. Do not install the sensor such that it is embedded in the heat source (such as in the barrel or heater) together with the lead pipe cover. This may heighten thermal conductivity from the ...

• If the pressure sensor is used together with the output converter

(Model PCT-300), always connect a grounding wire to the equipment side. (Ground the grounding terminal.) • Connect a grounding wire to the extruder. • For a heat-resistant glass covered cable, the cover is made of fibers. Therefore, the electrical insulation may deterior ...

6. LINEARIZING FUNCTION SETTING

„ Linearizing switch changing 1. First check the output characteristic type (A, N or B) engraved at the end of the figure showing the rated output value on the rating nameplate attached to the outer chamber of the pressure sensor. Example: Rating nameplate PRESSURE SENSOR TYPE CZ-100P-HB-SNE PRESSURE RANGE 0 - ...

* Main causes:

These causes may arise independently or in mutual relations. • Generative cause in operation Overpressure Load pressure exceeding its limit applied. Irregular thread engagement The metal pressure sensing surface mechanically scratched or chipped off. Metal fatigue The metal sensing surface fatigued by the application of changi ...

® RKC INSTRUMENT INC.

HEADQUARTERS: 16-6, KUGAHARA 5-CHOME, OHTA-KU TOKYO 146-8515 JAPAN PHONE: 03-3751-9799 (+81 3 3751 9799) E-mail: [email protected] FAX: 03-3751-8585 (+81 3 3751 8585) The first edition: APR. 1995 The second edition: MAY 2004 [IMQ00] Maximum diaphragm: 400 °C Maximum strain gauge temperature: 150 °C When the t ...