TSI Instruments Quest The Edge Quick Start Manual

Text of the manual

Edge Display and Keypad

Turning on 1. Turn on the Edge by pressing and releasing the key. Resetting Resetting (or clearing the data) is performed by either using the TSI Quest Detection Management Software DMS when downloading or setting up parameters. Another option is from the Edge’s shortcut reset menu. + To reset from the keypad ensure th

A calibration may be performed before and/or after a dosimetry test.

1. Remove the windscreen while the Edge is turned on and stopped. 2. Slide the Edge into the cavity of the calibrator, ensuring the adapter ring is inserted into the calibrator. 3. Turn on the calibrator. Set the calibrator to 114 dB and 1000 Hz (if applicable). 4. “CAL” w

Running a study

1. Once the Edge is turned on, press and hold the button. + A 3, 2, 1 countdown will appear. Then, a run indicator will appear on the top of the display. Also the LED will flash green unless this was disabled in the DMS via the heartbeat field. Ending a study 1. While running a study, press and hold the run/stop

When the battery is fully discharged, allow approximately 3 to 4 hours to recharge your Edge.

Tip: to quickly check battery power remaining on the Edge, press the up/down arrow key to view the battery power symbol and hours on the display. (See user manual for customizing display if it does not appear.) 1. To charge, plug in the power adapter to th

Downloading data and DMS

1. To communicate with DMS and the Edge, turn on the Edge and plug in the USB cable from the edge to the PC. Then, open TSI Quest Detection Management software DMS. 2. Click on the button. The instrument communications panel will appear. Select the Edge instrument and model from the dosimetry family. 3. C

Common noise dosimeter terms

Criterion Level (CL) - expressed in decibels (dB), it is the maximum allowable accumulated noise level that results in 100% dose. Regional noise standards specify criterion level. Dose - expressed in percent, it is the percent of the maximum exposure that has accumulated over the run time. 100% is the maximum allowable exposure.

Quest is a trademark, and TSI and the

TSI logo are registered trademarks of TSI Incorporated. TSI Incorporated - Visit our website www.tsi.com for more information. North America and Asia Pacific Telephone: +1 800-680-1220 (USA) +1 651-490-2860 (Outside USA) Fax: +1 651-490-3824 E-mail: [email protected] Europe, Middle East, and Africa Telephone: