KERN PLS Operating Instructions Manual
File Specifications
| Brand: | KERN |
| Category: | Scales |
| Model: | PLS (Operating Instructions Manual), prq03N8LwoxuU |
| File Info: | Adobe Acrobat PDF (DjVu) |
| File Size: | — |
| Pages: | 34 pages |
Download PDF
Download KERN PLS Manual (34 pages)Read Online:
Read the KERN PLS Operating Instructions Manual online using the built-in PDF viewer. This document explains how to operate the device, adjust settings and perform regular maintenance.
Key Pages of Document
6.5.3 Balance zero-display
If the balance does not show exactly zero although the pan scale is unloaded, press the TARE-button and the balance will be reset to zero [ a ]. 6.5.4 Auto-zero-function In the menu you can turn on or off the auto-zero-function. To do so, proceed as follows: Ö Turn on balance using the ON/OFF-button Ö Press the Tare-button; the d ...
3.4 Monitoring the test substances
KERN PLS The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the answerable user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN’s home page (ww ...
9 Data output RS 232 C
KERN PLS 9.1 Technical Data 8-bit ASCII code 8 Data bits, 1 stop bit, no parity bit Baud rate selectable for 300, 600, 1200, 2400, 4800 and 9600 Baud Miniature plug necessary (9 pol D-Sub) For operation with interface faultless operation is ensured only with the respective KERN- interface cable (max. 2m) 9.2 Pin allocation of the ...
6.6.1 Adjusting with internal weight (only PLJ/ALJ)
KERN PLS The installed adjusting weight allows checking and resetting the weighing accuracy at any time. The automatic adjustment function is always active. You can also start adjustment manually by pressing the Cal-button. Automatic adjustment is started when the balance is • cut off from power supply, • when c ...
Weighing range (Max)
360 g 510 g 2,100 g 4,000 g Taring range (subtractive) 360 g 510 g 2,100 g 4,000 g Reproducibility 0.002 g 0.002 g 0.02 g 0.02 g Linearity ± 0.002 g ± 0.002 g ± 0.02 g ± 0.02 g Minimal piece weight for Parts counting > 0.005 g > 0.005 g > 0.05 ...
9.4.3 Output format
KERN PLS Stable values: 1 2 3 S CR LF 1 2 3 4 5 6 S B A B CR LF Display 0.000 g 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 1 8 19 20 S B B B + B B B B 0 0 0 0 B B g B CR LF Display 45,288 g 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 1 8 19 20 S B B B + B B B 4 5 2 8 8 B B g B CR LF Instable values: 1 2 3 4 S I CR LF Display 0.000 g 1 2 3 4 5 6 7 8 9 ...
Weighing range (Max)
KERN PLS 360 g 510 g 2,100 g 4,000 g 6,100 g Minimum load (Min) 0,02 g 0,02 g 0,5 g 0,5 g - Taring range (subtractive) 360 g 510 g 2,100 g 4,000 g 6,100 g Reproducibility 0.002 g 0.002 g 0.02 g 0.02 g 0.01 g Linearity ± 0.002 g ± 0 ...
The balance should be switched off for a short time following an interruption in the
KERN PLS programme sequence and disconnected from the mains supply. It is then necessary to repeat the weighing process from the beginning. Help: Fault Possible cause Weight display is not illuminated. • The balance is not switched on ...
The dead weight of any type of weighing container can be tared out by pressing a
KERN PLS button, so that subsequent weighing procedures show the net weight of the items to be weighed. Operation: Display: Ö Use the ON/OFF-button to turn on the balance Ö As soon as the weight display shows „0.0000" your balance is ready for weighing 0.0000 g Ö Place i ...
KERN PLS user guide recommended for: PKS 200-3, HDB 5K5N, MBC-(E), CB, CXB.