LIMIT 500 Operating Manual
| Brand: | LIMIT |
| Category: | Multimeter |
| Model: | 500 (Operating Manual) |
| File Type: | |
| File Size: | — |
| Pages: | 14 |
Text of LIMIT 500 Operating Manual
LIMIT 500 Fig 1. Voltage measurement
DC and AC
Fig 2. Current measurement AC
Fig 3. Diode test
Continuity test
Resistance
Fig 4. Replacing battery
Fig 5. Replacing battery
500
LIMIT 500 Illustrations & Tables
1
DC Voltage
Range
Resolution
Accuracy
Overload
Protection
200mV
0.1mV
250V AC
2V
1mV
±(0,5%+1)
20V
10mV
1000V AC
200V
100mV
1000V
1V
±(0,8%+2)
AC Voltage
Range
Resolution
Accuracy
Overload
Protection
2V
10mV
20V
10V
±(0.8%+3)
1000V AC
200V
100V
1000V
1V
±(1.2%+3)
DC Current
Range
Resolution
Accuracy
Overload
Protection
2mA
1μA
±(0.8%+1)
CE Version:Fuse 0.5A,
200mA
0.1mA
±(1.5%+1)
250V, fast type, 5x20mm
20mA
10mA
±(2%+5)
Un-
LIMIT 500 CE Version: Fuse 0.5A,
200mA
0.1mA
±(1.8%+3)
250V, fast type, 5x20mm
20mA
10mA
±(3.0%+5)
Resistance
Range
Resolution
Accuracy
Overload
Protection
200Ω
0.1Ω
±(0.8%+3) + Test Lead
2Ω
1Ω
Short Circuit Resistence
250V AC
20kΩ
10Ω
±(0.8%+1)
2kΩ
1kΩ
20MΩ
10MΩ
±(1.0%+2)
Capacitance
Range
Resolution
Accuracy
2nF
1pF
±(4.0%+3)
200nF
0.1nF
100μF
0.1μF
±(5.0%+4) When it is
Accuracy ±(a%reading + b digits)
hFE
1ß
Vce≈3V Ibo≈10μA 1000ßMAX
Frequency (UT58C only)
Range
Resolution
Accuracy
Overload
Protection
2kHz
1Hz
20Hz
10Hz
± (1.5%+5)
250V AC
Remarks
• 100mVrms < input amplitude < 30Vrms
3
Tabels
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Svenska . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-19
Norsk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20-26
Dansk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27-34
Suomi . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35-41
Deutsch . . . . . . . . . . . . . . . . . . . . . . . . . . .
Voltage DC and AC
Current DC and AC
Resistance
Temperature
Diodes test
Continuity test
Capacitance
Transistor test
Frequency
Battery
Fuses
Contents
This Operating Manual covers information on safety and cautions.
Please read the relevant information carefully and observe all the
Warnings and Notes strictly.
Limit 500 are 3 1/2 digits instrument for professional use. Display
have large digits and also shows correct test leads terminals and
rotary switch position, makes this instrument easy to handle for the user.
General Specifications
Measuring range and accuracy see page 1-3.
•
Fused Prote