LIMIT Limit 21 fig.1 DC/AC Voltage Measurement

Red Black fig.2 Testing for Continuity 40A 400A fig.3 DC/AC Current Measurement Screw fig.4 Replacing the Battery Illustrations 1

LIMIT Limit 21 A. DC Voltage

Range Resolution Accuracy Overload protection 400.0mV 4.000V 40.00V 400.0V 600V 0.1mV 1mV 10mV 100mV 1V (0.8%+3) (0.8%+1) (1%+3) 600V DC/AC B. AC Voltage Range Resolution Accuracy Overload protection 4.000V 40.00V 400.0V 600V 1mV 10mV 100mV 1V (1%+5) (1.2%+5) 600V DC/AC 21

D. Diode Test

Range Resolution Accuracy Overload protection 1mV Display forward voltage drop nearest value 600Vp E. Continuity Test Range Resolution Accuracy Overload protection 100m Around 50 , the buzzer beeps 600Vp Tables

For reference only

G. Duty Cycle Range Resolution Accuracy Overload protection 0.1% For reference only 600Vp 0.1%~99.9% 4 21

H. DC Current

Range Resolution Accuracy Overload protection 0.01A 0.1A 400A DC/AC 40.00A 400.0A (2%+5) (2%+3) I. AC Current Range Resolution Accuracy Overload protection 0.01A 0.1A 400A DC/AC 40.00A 400.0A (2.5%+8) (2.5%+5) 50Hz ~ 60Hz Frequency Response Tables