Bante Instruments ECscan20 Instruction Manual
Text of the manual
Thank you for selecting the ECscan series pocket conductivity tester. This manual provides a step-by-step guide to help you operate the instrument,
please carefully read the following instructions according to the model you have purchased.
Installing the Batteries
1.
Twist the electrode collar counter clockwise, pull the electrode awa
Switching the Tester On and Off
•
Press and hold the Meas key for about 5 seconds to switch on the tester.
•
Press and hold the Meas key to switch off the tester.
•
To enable the auto-power off function, refer to the Setup Menu section.
Setup Menu
The ECscan series tester contains 7 menu items in the setup menu, the following table descr
The ECscan series conductivity tester allows 1 to 3 points calibration. For better accuracy, we recommend that you perform 3 points calibration
or select a standard solution closest to the sample conductivity you are measuring. The tester will automatically detect the calibration standard
and prompt the user to perform
The ECscan series tester has a built-in temperature sensor that used for automatic temperature compensation (ATC). Durning the measurement,
if the measured temperature reading differs from that of an accurate thermometer, the tester needs to be calibrated.
1.
Press and hold the Cal key to enter the setup menu.
2.
Press
Preparation of Conductivity Standard Solutions
1.
Place the analytical grade potassium chloride (KCl) reagent in a beaker and dry in an oven for about 3 hours at 105°C/221°F, then cool to
room temperature.
2.
Add the reagent to a 1 litre volumetric flask according to the instructions in table below.
3.
Fill the deionized water to th
0 to 20.00, 200.0, 2000 μS/cm, 20.00 mS/cm
Resolution
0.01, 0.1, 1
Accuracy
±1% F.S.
Calibration Point
1, 2 or 3 points
Calibration Solution
84 μS/cm, 1413 μS/cm, 12.88 mS/cm
TDS
Range
---
0 to 10.00, 100.0, 1000 ppm, 20.00 ppt
Resolution
---
0.01, 0.1, 1
Accuracy
---
±1%
Bante Instruments Inc.
Office: 4715 Castlewood St., Sugar land, TX 77479, USA
Tel: (+1) 346-762-7358
E-mail: [email protected]
Factory: A10, No.2066, Laifang Rd., Shanghai 201615, China
Tel: (+86) 21-6404-1598
E-mail: [email protected]