Bante Instruments ECscan20 Instruction Manual

Text of the manual

Thank you for selecting the ECscan series pocket conductivity tester. This manual provides a step-by-step guide to help you operate the instrument,

please carefully read the following instructions according to the model you have purchased. Installing the Batteries 1. Twist the electrode collar counter clockwise, pull the electrode awa

Switching the Tester On and Off

• Press and hold the Meas key for about 5 seconds to switch on the tester. • Press and hold the Meas key to switch off the tester. • To enable the auto-power off function, refer to the Setup Menu section. Setup Menu The ECscan series tester contains 7 menu items in the setup menu, the following table descr

The ECscan series conductivity tester allows 1 to 3 points calibration. For better accuracy, we recommend that you perform 3 points calibration

or select a standard solution closest to the sample conductivity you are measuring. The tester will automatically detect the calibration standard and prompt the user to perform

The ECscan series tester has a built-in temperature sensor that used for automatic temperature compensation (ATC). Durning the measurement,

if the measured temperature reading differs from that of an accurate thermometer, the tester needs to be calibrated. 1. Press and hold the Cal key to enter the setup menu. 2. Press

Preparation of Conductivity Standard Solutions

1. Place the analytical grade potassium chloride (KCl) reagent in a beaker and dry in an oven for about 3 hours at 105°C/221°F, then cool to room temperature. 2. Add the reagent to a 1 litre volumetric flask according to the instructions in table below. 3. Fill the deionized water to th

0 to 20.00, 200.0, 2000 μS/cm, 20.00 mS/cm

Resolution 0.01, 0.1, 1 Accuracy ±1% F.S. Calibration Point 1, 2 or 3 points Calibration Solution 84 μS/cm, 1413 μS/cm, 12.88 mS/cm TDS Range --- 0 to 10.00, 100.0, 1000 ppm, 20.00 ppt Resolution --- 0.01, 0.1, 1 Accuracy --- ±1%

Bante Instruments Inc.

Office: 4715 Castlewood St., Sugar land, TX 77479, USA Tel: (+1) 346-762-7358 E-mail: [email protected] Factory: A10, No.2066, Laifang Rd., Shanghai 201615, China Tel: (+86) 21-6404-1598 E-mail: [email protected]  www.bante-china.com

More Bante Instruments Test Equipment Owners Instructions: